» Articles » PMID: 23143809

Reduced Administered Activity, Reduced Acquisition Time, and Preserved Image Quality for the New CZT Camera

Overview
Journal J Nucl Cardiol
Date 2012 Nov 13
PMID 23143809
Citations 30
Authors
Affiliations
Soon will be listed here.
Abstract

Background: For a 1-day myocardial perfusion SPECT (MPS) the recommendations for administered activity stated in the EANM guidelines results in an effective dose of up to 16 mSv per patient. Recently, a gamma camera system, based on cadmium zinc telluride (CZT) technology, was introduced. This technique has the potential to reduce the effective dose and scan time compared to the conventional NaI gamma camera. The aim of this study was to investigate if the effective dose can be reduced with a preserved image quality using CZT technology in MPS.

Methods: In total, 150 patients were included in the study. All underwent a 1-day (99m)Tc-tetrofosmin stress-rest protocol and were divided into three subgroups (n = 50 in each group) with 4, 3, and 2.5 MBq/kg body weight of administered activity in the stress examination, respectively. The acquisition time was increased in proportion to the decrease in administered activity. All examinations were analyzed for image quality by visual grading on a 4-point scale (1 = poor, 2 = adequate, 3 = good, 4 = excellent), by two expert readers.

Results: The total effective dose (stress + rest) decreased from 9.3 to 5.8 mSv comparing 4 to 2.5 MBq/kg body weight. For the patients undergoing stress examination only (35%) the effective dose, administrating 2.5 MBq/kg, was 1.4 mSv. The image acquisition times for 2.5 MBq/kg body weight were 475 and 300 seconds (stress and rest) compared to 900 seconds for each when using conventional MPS. The average image quality was 3.7 ± 0.5, 3.8 ± 0.5, and 3.8 ± 0.4 for the stress images and 3.5 ± 0.6, 3.6 ± 0.6, and 3.5 ± 0.6 for the rest images and showed no statistically significant difference (P = .62) among the 4, 3, and 2.5 MBq/kg groups.

Conclusions: The new CZT technology can be used to considerably decrease the effective dose and acquisition time for MPS with preserved high image quality.

Citing Articles

Performance evaluation of the 3D-ring cadmium-zinc-telluride (CZT) StarGuide system according to the NEMA NU 1-2018 standard.

Zorz A, Rossato M, Turco P, Colombo Gomez L, Bettinelli A, De Monte F EJNMMI Phys. 2024; 11(1):69.

PMID: 39052176 PMC: 11272762. DOI: 10.1186/s40658-024-00671-x.


.

Sweet C, Shmuel N, Shoaf J, Stoecklein M, Muthukrishnan A, Stern E Nucl Med Mol Imaging. 2024; 58(1):1-8.

PMID: 38250182 PMC: 10796310. DOI: 10.1007/s13139-023-00825-2.


Optimization of Tl Dosage in a Simultaneous Acquisition Rest Tc/Stress Tl Dual-Isotope Myocardial Perfusion Single-Photon Emission Computed Tomography with Semiconductor Gamma Camera.

Sugai S, Matsumoto N, Makita A, Kuronuma K, Suzuki Y, Hori Y Ann Nucl Cardiol. 2023; 9(1):68-72.

PMID: 38058585 PMC: 10696141. DOI: 10.17996/anc.23-00010.


Immediate recruitment of dormant coronary collaterals can provide more than half of normal resting perfusion during coronary occlusion in patients with coronary artery disease.

Reid B, Lindow T, Warren S, Persson E, Bhindi R, Ringborn M J Nucl Cardiol. 2023; 30(6):2338-2345.

PMID: 37280387 PMC: 10682227. DOI: 10.1007/s12350-023-03271-x.


Same same but different: dopamine transporter SPECT on scanners with CZT vs. NaI detectors.

Thiele F, Schau F, Rogasch J, Wetz C, Bluemel S, Brenner W EJNMMI Res. 2023; 13(1):24.

PMID: 36949290 PMC: 10033816. DOI: 10.1186/s13550-023-00973-8.


References
1.
Cerqueira M, Allman K, Ficaro E, Hansen C, Nichols K, Thompson R . Recommendations for reducing radiation exposure in myocardial perfusion imaging. J Nucl Cardiol. 2010; 17(4):709-18. DOI: 10.1007/s12350-010-9244-0. View

2.
Duvall W, Croft L, Ginsberg E, Einstein A, Guma K, George T . Reduced isotope dose and imaging time with a high-efficiency CZT SPECT camera. J Nucl Cardiol. 2011; 18(5):847-57. DOI: 10.1007/s12350-011-9379-7. View

3.
Eisenberg M, Afilalo J, Lawler P, Abrahamowicz M, Richard H, Pilote L . Cancer risk related to low-dose ionizing radiation from cardiac imaging in patients after acute myocardial infarction. CMAJ. 2011; 183(4):430-6. PMC: 3050947. DOI: 10.1503/cmaj.100463. View

4.
Hesse B, Tagil K, Cuocolo A, Anagnostopoulos C, Bardies M, Bax J . EANM/ESC procedural guidelines for myocardial perfusion imaging in nuclear cardiology. Eur J Nucl Med Mol Imaging. 2005; 32(7):855-97. DOI: 10.1007/s00259-005-1779-y. View

5.
Buechel R, Herzog B, Husmann L, Burger I, Pazhenkottil A, Treyer V . Ultrafast nuclear myocardial perfusion imaging on a new gamma camera with semiconductor detector technique: first clinical validation. Eur J Nucl Med Mol Imaging. 2010; 37(4):773-8. DOI: 10.1007/s00259-009-1375-7. View