» Articles » PMID: 19339672

A Novel High-sensitivity Rapid-acquisition Single-photon Cardiac Imaging Camera

Overview
Journal J Nucl Med
Specialty Nuclear Medicine
Date 2009 Apr 3
PMID 19339672
Citations 108
Authors
Affiliations
Soon will be listed here.
Abstract

Unlabelled: This study described and validated a new solid-state single-photon gamma-camera and compared it with a conventional-SPECT Anger camera. The compact new camera uses a unique method for localizing gamma-photon information with a bank of 9 solid-state detector columns with tungsten collimators that rotate independently.

Methods: Several phantom studies were performed comparing the new technology with conventional-SPECT technology. These included measurements of line sources and single- and dual-radionuclide studies of a torso phantom. Simulations were also performed using a cardiothoracic phantom. Furthermore, 18 patients were scanned with both the new camera and a conventional-SPECT camera.

Results: The new camera had a count sensitivity that was 10 times higher than that of the conventional camera and a compensated spatial resolution that was moderately better. Dual-radionuclide studies using a phantom show the further potential of the new camera for a 2-tracer simultaneous acquisition. Two-minute clinical studies with the new camera and 11-min studies with the conventional camera qualitatively showed good-to-excellent image quality and improved myocardial edge definition for the new camera.

Conclusion: These initial performance characteristics of a new solid-state single-photon gamma-camera offer great promise for clinical dynamic SPECT protocols, with important implications for applications in nuclear cardiology and molecular imaging.

Citing Articles

A review of state-of-the-art resolution improvement techniques in SPECT imaging.

Cheng Z, Chen P, Yan J EJNMMI Phys. 2025; 12(1):9.

PMID: 39883257 PMC: 11782768. DOI: 10.1186/s40658-025-00724-9.


Recent Progress in Myocardial Perfusion Imaging Techniques.

Jo K, Kang W Nucl Med Mol Imaging. 2024; 58(7):400-405.

PMID: 39635632 PMC: 11612050. DOI: 10.1007/s13139-024-00852-7.


Optimization and application of renal depth measurement method in the cadmium-zinc-telluride‑based SPECT/CT renal dynamic imaging.

Zheng H, Li X, Wang S, Hou S, Shi C, Li X EJNMMI Phys. 2024; 11(1):96.

PMID: 39542940 PMC: 11564595. DOI: 10.1186/s40658-024-00702-7.


Technological Advances in SPECT and SPECT/CT Imaging.

Bouchareb Y, AlSaadi A, Zabah J, Jain A, Al-Jabri A, Phiri P Diagnostics (Basel). 2024; 14(13).

PMID: 39001321 PMC: 11241697. DOI: 10.3390/diagnostics14131431.


Impact of cardiac size on diagnostic performance of single-photon emission computed tomography myocardial perfusion imaging: insights from the REgistry of Fast Myocardial Perfusion Imaging with NExt generation single-photon emission computed....

Randazzo M, Elias P, Poterucha T, Sharir T, Fish M, Ruddy T Eur Heart J Cardiovasc Imaging. 2024; 25(7):996-1006.

PMID: 38445511 PMC: 11210974. DOI: 10.1093/ehjci/jeae055.