Purpose: To assess the diagnostic performance of a novel ultrafast cardiac gamma camera with cadmium-zinc-telluride (CZT) solid-state semiconductor detectors for nuclear myocardial perfusion imaging (MPI).

Methods: The study group comprised 75 consecutive patients (55 men, BMI range 19-45 kg/m(2)) who underwent a 1-day (99m)Tc-tetrofosmin adenosine-stress/rest imaging protocol. Scanning was performed first on a conventional dual-detector SPECT gamma camera (Ventri, GE Healthcare) with a 15-min acquisition time each for stress and rest. All scans were immediately repeated on an ultrafast CZT camera (Discovery 530 NMc, GE Healthcare) with a 3-min scan time for stress and a 2-min scan time for rest. Clinical agreement (normal, ischaemia, scar) between CZT and SPECT was assessed for each patient and for each coronary territory using SPECT MPI as the reference standard. Segmental myocardial tracer uptake values (percent of maximum) using a 20-segment model and left ventricular ejection fraction (EF) values obtained using CZT were compared with those obtained using conventional SPECT by intraclass correlation and by calculating Bland-Altman limits of agreement.

Results: There was excellent clinical agreement between CZT and conventional SPECT on a per-patient basis (96.0%) and on a per-vessel territory basis (96.4%) as shown by a highly significant correlation between segmental tracer uptake values (r=0.901, p<0.001). Similarly, EF values for both scanners were highly correlated (r=0.976, p<0.001) with narrow Bland-Altman limits of agreement (-5.5-10.6%).

Conclusion: The novel CZT camera allows a more than fivefold reduction in scan time and provides clinical information equivalent to conventional standard SPECT MPI.

Citing Articles

New Generation SPECT Cameras Based on Cadmium-Zinc Telluriide Technology.

Matsunari I Ann Nucl Cardiol. 2024; 10(1):59-63.

PMID: 39635329 PMC: 11612397. DOI: 10.17996/anc.24-00011.


Technological Advances in SPECT and SPECT/CT Imaging.

Bouchareb Y, AlSaadi A, Zabah J, Jain A, Al-Jabri A, Phiri P Diagnostics (Basel). 2024; 14(13).

PMID: 39001321 PMC: 11241697. DOI: 10.3390/diagnostics14131431.


Same same but different: dopamine transporter SPECT on scanners with CZT vs. NaI detectors.

Thiele F, Schau F, Rogasch J, Wetz C, Bluemel S, Brenner W EJNMMI Res. 2023; 13(1):24.

PMID: 36949290 PMC: 10033816. DOI: 10.1186/s13550-023-00973-8.


The Effect of an Energy Window with an Ellipsoid Phantom on the Differential Defect Contrast on Myocardial SPECT Images.

Oglat A, Sayah M Bioengineering (Basel). 2022; 9(8).

PMID: 35892754 PMC: 9331383. DOI: 10.3390/bioengineering9080341.


Diagnostic Value of Single-Photon Emission Tomography Stress Test in Patients With Suspected Coronary Artery Disease in Saudi Arabia.

Alqarni M, Bukhari Z, Abukhodair A, Binammar D, Alzahrani A, Alkahtani A Cureus. 2021; 13(10):e19071.

PMID: 34824946 PMC: 8610429. DOI: 10.7759/cureus.19071.


References
1.
Valenta I, Treyer V, Husmann L, Gaemperli O, Schindler M, Herzog B . New reconstruction algorithm allows shortened acquisition time for myocardial perfusion SPECT. Eur J Nucl Med Mol Imaging. 2009; 37(4):750-7. DOI: 10.1007/s00259-009-1300-0. View

2.
Rybicki F, Otero H, Steigner M, Vorobiof G, Nallamshetty L, Mitsouras D . Initial evaluation of coronary images from 320-detector row computed tomography. Int J Cardiovasc Imaging. 2008; 24(5):535-46. DOI: 10.1007/s10554-008-9308-2. View

3.
Kapur A, Latus K, Davies G, Dhawan R, Eastick S, Jarritt P . A comparison of three radionuclide myocardial perfusion tracers in clinical practice: the ROBUST study. Eur J Nucl Med Mol Imaging. 2002; 29(12):1608-16. DOI: 10.1007/s00259-002-0998-8. View

4.
Thomas G, Miyamoto M, Morello 3rd A, Majmundar H, Thomas J, Sampson C . Technetium 99m sestamibi myocardial perfusion imaging predicts clinical outcome in the community outpatient setting. The Nuclear Utility in the Community (NUC) Study. J Am Coll Cardiol. 2004; 43(2):213-23. DOI: 10.1016/j.jacc.2003.07.041. View

5.
Giorgetti A, Rossi M, Stanislao M, Valle G, Bertolaccini P, Maneschi A . Feasibility and diagnostic accuracy of a gated SPECT early-imaging protocol: a multicenter study of the Myoview Imaging Optimization Group. J Nucl Med. 2007; 48(10):1670-5. DOI: 10.2967/jnumed.106.039107. View