Support Line and Tube Visibility in Chest Examinations Using Computed Radiography

Overview
Journal J Digit Imaging
Publisher Springer
Date 1997 Aug 1
PMID 9268908
Authors
Affiliations
Soon will be listed here.
Abstract

The visibility of support lines and tubes was compared in computed radiography (CR) and screen-film adult chest radiographs. Parameters investigated were radiation dose, image minification, and the use of unsharp mask enhancement. Five radiologists rated the visibility of support lines and tubes on a five-point scale ranging from 1 (entire course of line visible) to 5 (line not visible or only small portion seen). These CR results were compared with the visibility of support lines and tubes as assessed by the same readers for conventional screen-film radiographs (600 speed). Support line and tube visibility improved with image enhancement, image minification, and increasing radiation dose. At the same radiation exposure, support line and tube visibility of the screen-film combination was superior to that of standard CR images. Application of an unsharp mask enhancement algorithm to CR images, however, significantly improved support line and tube visibility in comparison to that obtained with the screen-film combination. It was concluded that unsharp mask processing is a valuable tool for improving the visibility of support lines and tubes in CR chest radiographs.

References
1.
Sonoda M, Takano M, Miyahara J, Kato H . Computed radiography utilizing scanning laser stimulated luminescence. Radiology. 1983; 148(3):833-8. DOI: 10.1148/radiology.148.3.6878707. View

2.
Huda W, Honeyman J, Frost M, Staab E . A cost-analysis of computed radiography and picture archiving and communication systems in portable radiography. J Digit Imaging. 1996; 9(1):39-44. DOI: 10.1007/BF03168567. View

3.
Fujita H, Ueda K, Morishita J, Fujikawa T, Ohtsuka A, Sai T . Basic imaging properties of a computed radiographic system with photostimulable phosphors. Med Phys. 1989; 16(1):52-9. DOI: 10.1118/1.596402. View

4.
Oestmann J, Greene R, Rubens J, Hall D, Robertson C, Llewellyn H . High frequency edge enhancement in the detection of fine pulmonary lines. Parity between storage phosphor digital images and conventional chest radiography. Invest Radiol. 1989; 24(9):643-6. DOI: 10.1097/00004424-198909000-00001. View

5.
Schaefer C, Greene R, Oestmann J, Kamalsky J, Hall D, Llewellyn H . Improved control of image optical density with low-dose digital and conventional radiography in bedside imaging. Radiology. 1989; 173(3):713-6. DOI: 10.1148/radiology.173.3.2813775. View