Basic Imaging Properties of a Computed Radiographic System with Photostimulable Phosphors

Overview
Journal Med Phys
Specialty Biophysics
Date 1989 Jan 1
PMID 2921980
Citations 13
Authors
Affiliations
Soon will be listed here.
Abstract

We measured the characteristic curve, modulation transfer function (MTF), and the Wiener spectrum of a commercially available computed radiographic (CR) system with photostimulable phosphor plate (imaging plate, IP). The characteristic curve (system response) obtained by an inverse-square x-ray sensitometry showed a wide dynamic range (order of 10(3) in maximum). The slit technique was employed to determine the MTF's, such as IP MTF, presampling MTF including the unsharpness of the detector (IP) and the blurring effect of the sampling aperture, and laser-printer MTF. It was found that the MTF of the standard type of IP was comparable to that of medium-speed screen/film systems. The noticeable degradation of resolution in our CR system, however, occurred at the stage of image data sampling: the presampling MTF was inferior to the IP MTF due to the effect of the scattering and resultant spreading of the incidence laser beam and the emitted luminescence. The noise was characterized by means of digital Wiener spectrum using uniformly exposed noise data. Exposure ranges could be separated into different sections depending upon the noise sources, such as quantum mottle at low exposure and system structure noise at high exposure.

Citing Articles

Effect of beam quality and readout direction in the edge profile on the modulation transfer function of photostimulable phosphor systems via the edge method.

Takarabe S, Kuramoto T, Shibayama Y, Tsuru H, Tatsumi M, Kato T J Med Imaging (Bellingham). 2021; 8(4):043501.

PMID: 34307736 PMC: 8291077. DOI: 10.1117/1.JMI.8.4.043501.


Development and Evaluation of the Fourier Spectral Distortion Metric.

Vent T, Acciavatti R, Maidment A IEEE Trans Med Imaging. 2020; 40(3):1055-1064.

PMID: 33326378 PMC: 10614697. DOI: 10.1109/TMI.2020.3045325.


Effect of differences in pixel size on image characteristics of digital intraoral radiographic systems: a physical and visual evaluation.

Kuramoto T, Takarabe S, Okamura K, Shiotsuki K, Shibayama Y, Tsuru H Dentomaxillofac Radiol. 2020; 49(6):20190378.

PMID: 32302229 PMC: 7461739. DOI: 10.1259/dmfr.20190378.


Application of a variable filter for presampled modulation transfer function analysis with the edge method.

Higashide R, Ichikawa K, Kunitomo H, Ohashi K Radiol Phys Technol. 2015; 8(2):320-30.

PMID: 26088943 DOI: 10.1007/s12194-015-0325-2.


Basic imaging properties of an indirect flat-panel detector system employing irradiation side sampling (ISS) technology for chest radiography: comparison with a computed radiographic system.

Tanaka N, Yano Y, Yabuuchi H, Akasaka T, Sasaki M, Ohki M Radiol Phys Technol. 2012; 6(1):162-9.

PMID: 23143711 DOI: 10.1007/s12194-012-0184-z.