Design and Realization of a Multi-coil Array for B Field Control in a Compact 1.5T Head-only MRI Scanner

Overview
Journal Magn Reson Med
Publisher Wiley
Specialty Radiology
Date 2023 May 5
PMID 37145035
Authors
Affiliations
Soon will be listed here.
Abstract

Purpose: To design and implement a multi-coil (MC) array for B field generation for image encoding and simultaneous advanced shimming in a novel 1.5T head-only MRI scanner.

Methods: A 31-channel MC array was designed following the unique constraints of this scanner design: The vertically oriented magnet is very short, stopping shortly above the shoulders of a sitting subject, and includes a window for the subject to see through. Key characteristics of the MC hardware, the B field generation capabilities, and thermal behavior, were optimized in simulations prior to its construction. The unit was characterized via bench testing. B field generation capabilities were validated on a human 4T MR scanner by analysis of experimental B fields and by comparing images for several MRI sequences acquired with the MC array to those acquired with the system's linear gradients.

Results: The MC system was designed to produce a multitude of linear and nonlinear magnetic fields including linear gradients of up to 10 kHz/cm (23.5 mT/m) with MC currents of 5 A per channel. With water cooling it can be driven with a duty cycle of up to 74% and ramp times of 500 μs. MR imaging experiments encoded with the developed multi-coil hardware were largely artifact-free; residual imperfections were predictable, and correctable.

Conclusion: The presented compact multi-coil array is capable of generating image encoding fields with amplitudes and quality comparable to clinical systems at very high duty cycles, while additionally enabling high-order B shimming capabilities and the potential for nonlinear encoding fields.

Citing Articles

High-quality lipid suppression and B0 shimming for human brain H MRSI.

Kumaragamage C, McIntyre S, Nixon T, De Feyter H, de Graaf R Neuroimage. 2024; 300:120845.

PMID: 39276817 PMC: 11540284. DOI: 10.1016/j.neuroimage.2024.120845.


Advancements in MR hardware systems and magnetic field control: B shimming, RF coils, and gradient techniques for enhancing magnetic resonance imaging and spectroscopy.

Shang Y, Simegn G, Gillen K, Yang H, Han H Psychoradiology. 2024; 4:kkae013.

PMID: 39258223 PMC: 11384915. DOI: 10.1093/psyrad/kkae013.

References
1.
Sanchez H, Liu F, Trakic A, Weber E, Crozier S . 3D-gradient coil structures for MRI designed using fuzzy membership functions. Annu Int Conf IEEE Eng Med Biol Soc. 2007; 2007:2042-5. DOI: 10.1109/IEMBS.2007.4352721. View

2.
Zur Y . Design of improved spectral-spatial pulses for routine clinical use. Magn Reson Med. 2000; 43(3):410-20. DOI: 10.1002/(sici)1522-2594(200003)43:3<410::aid-mrm13>3.0.co;2-3. View

3.
Stockmann J, Aksit Ciris P, Galiana G, Tam L, Constable R . O-space imaging: Highly efficient parallel imaging using second-order nonlinear fields as encoding gradients with no phase encoding. Magn Reson Med. 2010; 64(2):447-56. PMC: 3147299. DOI: 10.1002/mrm.22425. View

4.
Romeo F, Hoult D . Magnet field profiling: analysis and correcting coil design. Magn Reson Med. 1984; 1(1):44-65. DOI: 10.1002/mrm.1910010107. View

5.
Snyder A, Corum C, Moeller S, Powell N, Garwood M . MRI by steering resonance through space. Magn Reson Med. 2013; 72(1):49-58. PMC: 4285333. DOI: 10.1002/mrm.24888. View