» Articles » PMID: 23913527

MRI by Steering Resonance Through Space

Overview
Journal Magn Reson Med
Publisher Wiley
Specialty Radiology
Date 2013 Aug 6
PMID 23913527
Citations 11
Authors
Affiliations
Soon will be listed here.
Abstract

Purpose: This work introduces a technique to excite MR signals locally and to steer this localized region over the object in a spatiotemporal manner. The purpose is to demonstrate the feasibility of MRI with multidimensional spatiotemporal-encoding in a way that provides the ability to compensate extreme field inhomogeneity.

Methods: The method is called steering resonance over the object (STEREO). A modulated gradient is applied in concert with a frequency-modulated pulse to steer a resonant region through space and thus produce sequential excitation and echo formation. Images are reconstructed using exclusively an inverse problem solution.

Results: Images of phantoms and human brain were produced to demonstrate the feasibility of the STEREO sequence and image reconstruction. Simulations support the postulated capability to compensate for extreme field inhomogeneity.

Conclusion: STEREO represents a substantial departure from conventional MRI in which spins contained in the sample, slab, or slice are excited synchronously. By exciting spins sequentially along a curved spatial trajectory, STEREO in principle affords a unique opportunity to adjust for spatial variations in static and radiofrequency fields. By adjusting field amplitudes and frequencies in a temporal manner in STEREO, in future works it should be possible to perform MRI with highly inhomogeneous fields.

Citing Articles

Advancements in MR hardware systems and magnetic field control: B shimming, RF coils, and gradient techniques for enhancing magnetic resonance imaging and spectroscopy.

Shang Y, Simegn G, Gillen K, Yang H, Han H Psychoradiology. 2024; 4:kkae013.

PMID: 39258223 PMC: 11384915. DOI: 10.1093/psyrad/kkae013.


Design and realization of a multi-coil array for B field control in a compact 1.5T head-only MRI scanner.

Theilenberg S, Shang Y, Ghazouani J, Kumaragamage C, Nixon T, McIntyre S Magn Reson Med. 2023; 90(3):1228-1241.

PMID: 37145035 PMC: 10330274. DOI: 10.1002/mrm.29692.


Reducing the Complexity of Model-Based MRI Reconstructions via Sparsification.

Gutierrez A, Mullen M, Xiao D, Jang A, Froelich T, Garwood M IEEE Trans Med Imaging. 2021; 40(9):2477-2486.

PMID: 33999816 PMC: 8569912. DOI: 10.1109/TMI.2021.3081013.


Design of an Intraoral Dipole Antenna for Dental Applications.

Ozen A, Idiyatullin D, Adriany G, Jungst S, Kobayashi N, Groenke B IEEE Trans Biomed Eng. 2021; 68(8):2563-2573.

PMID: 33513097 PMC: 9720844. DOI: 10.1109/TBME.2021.3055777.


Dynamic multicoil technique (DYNAMITE) MRI on human brain.

Juchem C, Theilenberg S, Kumaragamage C, Mullen M, DelaBarre L, Adriany G Magn Reson Med. 2020; 84(6):2953-2963.

PMID: 32544274 PMC: 8168279. DOI: 10.1002/mrm.28323.


References
1.
Ben-Eliezer N, Irani M, Frydman L . Super-resolved spatially encoded single-scan 2D MRI. Magn Reson Med. 2010; 63(6):1594-600. DOI: 10.1002/mrm.22377. View

2.
Idiyatullin D, Corum C, Park J, Garwood M . Fast and quiet MRI using a swept radiofrequency. J Magn Reson. 2006; 181(2):342-9. DOI: 10.1016/j.jmr.2006.05.014. View

3.
Goerke U, Garwood M, Ugurbil K . Functional magnetic resonance imaging using RASER. Neuroimage. 2010; 54(1):350-60. PMC: 2975516. DOI: 10.1016/j.neuroimage.2010.08.011. View

4.
Vaughan J, Hetherington H, Otu J, Pan J, Pohost G . High frequency volume coils for clinical NMR imaging and spectroscopy. Magn Reson Med. 1994; 32(2):206-18. DOI: 10.1002/mrm.1910320209. View

5.
Van de Walle R, Barrett H, Myers K, Altbach M, Desplanques B, Gmitro A . Reconstruction of MR images from data acquired on a general nonregular grid by pseudoinverse calculation. IEEE Trans Med Imaging. 2001; 19(12):1160-7. DOI: 10.1109/42.897806. View