Phase-stable Doppler OCT at 19 MHz Using a Stretched-pulse Mode-locked Laser

Overview
Specialty Radiology
Date 2018 Mar 16
PMID 29541496
Citations 11
Authors
Affiliations
Soon will be listed here.
Abstract

We present a swept-wavelength optical coherence tomography (OCT) system with a 19 MHz laser source and electronic phase-locking of the source, acquisition clock, and beam scanning mirrors. The laser is based on stretched-pulse active mode-locking using an electro-optic modulator. Beam scanning in the fast axis uses a resonant micro-electromechanical systems (MEMS) -based mirror at ~23.8 kHz. Acquisition is performed at 1.78 Gigasamples per second using an external fixed clock. Phase sensitive imaging without need for k-clocking, A-line triggers, or phase-calibration methods is demonstrated. The system was used to demonstrate inter-frame and inter-volume Doppler imaging in the mouse ear and brain at 4D acquisition rates of 1, 30, 60 and 100 volumes/sec (V-scans/s). Angiography based on inter-frame and inter-volume methods are presented. The platform offers extremely fast and phase-stable measurements that can be used in preclinical angiographic and Doppler investigations of perfusion dynamics.

Citing Articles

Optical coherence tomography.

Bouma B, de Boer J, Huang D, Jang I, Yonetsu T, Leggett C Nat Rev Methods Primers. 2023; 2.

PMID: 36751306 PMC: 9901537. DOI: 10.1038/s43586-022-00162-2.


Wide-Field Three-Dimensional Depth-Invariant Cellular-Resolution Imaging of the Human Retina.

Lee B, Jeong S, Lee J, Kim T, Braaf B, Vakoc B Small. 2023; 19(11):e2203357.

PMID: 36642824 PMC: 10023497. DOI: 10.1002/smll.202203357.


Investigation of multiple scattering in space and spatial-frequency domains: with application to the analysis of aberration-diverse optical coherence tomography.

Wu M, Liu S, Leartprapun N, Adie S Biomed Opt Express. 2022; 12(12):7478-7499.

PMID: 35003847 PMC: 8713691. DOI: 10.1364/BOE.439395.


Robust and easy-to-operate stretched-pulse mode-locked wavelength-swept laser with an all-polarization-maintaining fiber cavity for 10  MHz A-line rate optical coherence tomography.

Joo J, Kim T, Vakoc B, Oh W Opt Lett. 2021; 46(16):3857-3860.

PMID: 34388759 PMC: 8455078. DOI: 10.1364/OL.424835.


Correction of circumferential and longitudinal motion distortion in high-speed catheter/endoscope-based optical coherence tomography.

Nguyen T, Ahsen O, Liang K, Zhang J, Mashimo H, Fujimoto J Biomed Opt Express. 2021; 12(1):226-246.

PMID: 33520383 PMC: 7818954. DOI: 10.1364/BOE.409074.


References
1.
Wei X, Xu J, Xu Y, Yu L, Xu J, Li B . Breathing laser as an inertia-free swept source for high-quality ultrafast optical bioimaging. Opt Lett. 2014; 39(23):6593-6. DOI: 10.1364/OL.39.006593. View

2.
Wieser W, Biedermann B, Klein T, Eigenwillig C, Huber R . Multi-megahertz OCT: High quality 3D imaging at 20 million A-scans and 4.5 GVoxels per second. Opt Express. 2010; 18(14):14685-704. DOI: 10.1364/OE.18.014685. View

3.
White B, Pierce M, Nassif N, Cense B, Park B, Tearney G . In vivo dynamic human retinal blood flow imaging using ultra-high-speed spectral domain optical coherence tomography. Opt Express. 2009; 11(25):3490-7. DOI: 10.1364/oe.11.003490. View

4.
Fingler J, Schwartz D, Yang C, Fraser S . Mobility and transverse flow visualization using phase variance contrast with spectral domain optical coherence tomography. Opt Express. 2009; 15(20):12636-53. DOI: 10.1364/oe.15.012636. View

5.
Wang S, Singh M, Lopez 3rd A, Wu C, Raghunathan R, Schill A . Direct four-dimensional structural and functional imaging of cardiovascular dynamics in mouse embryos with 1.5 MHz optical coherence tomography. Opt Lett. 2015; 40(20):4791-4. PMC: 4849121. DOI: 10.1364/OL.40.004791. View