Importance of Correct Patient Positioning in Myocardial Perfusion SPECT when Using a CZT Camera

Overview
Journal J Nucl Cardiol
Date 2014 May 9
PMID 24807623
Citations 18
Authors
Affiliations
Soon will be listed here.
Abstract

Introduction: Myocardial perfusion single photon emission computed tomography (MPS) is one of the most widely used diagnostic methods in patients with suspected ischemic heart disease (IHD). Recently, a novel technique based on cadmium-zinc-telluride (CZT) detectors, pinhole collimators, and a stationary gantry was introduced for MPS. The aim of this work was to investigate how patient positioning affects the reconstructed MPS images using this novel technique.

Materials And Methods: Eighteen patients referred for a clinical MPS due to suspected IHD were included in the study. All patients underwent MPS imaging on a GE Discovery NM 530c CZT camera. After image acquisition with the heart positioned in the center of the quality field of view (QFOV), the patients were re-imaged in different positions 5-20 mm off-center. The heart was still positioned within the limits of the QFOV during the off-center scans. The summed stress score and/or the summed rest score (SSS and/or SRS) for the acquisition performed in the center was compared to the same parameter for the acquisitions performed off-center.

Results: There was a statistically significant increase in SSS and/or SRS when imaging was performed with the heart 5-20 mm outside the center of the QFOV compared to optimal positioning (7.7 ± 1.3 vs 6.6 ± 1.3, P = .006). The SSS and/or SRS increased with ≥2 U in 35% (14/40) of the off-center examinations.

Conclusion: It is important to carefully position the patient's heart within the center of the QFOV when performing MPS with the Discovery NM 530c CZT camera to avoid positioning-related image artifacts that could affect the diagnostic accuracy.

Citing Articles

Improving multi-pinhole CZT myocardial perfusion imaging specificity without changing sensibility by using adapted filter parameters.

Vauchot F, DuBois J, Bourdon A EJNMMI Res. 2024; 14(1):27.

PMID: 38453724 PMC: 10920588. DOI: 10.1186/s13550-024-01083-9.


Performance evaluation of a novel multi-pinhole collimator on triple-NaI-detector SPECT/CT for dedicated myocardial imaging.

Krizsan A, Kukuts K, Al-Muhanna W, Szoboszlai Z, Balazs L, Szabo B EJNMMI Phys. 2023; 10(1):24.

PMID: 36964406 PMC: 10039219. DOI: 10.1186/s40658-023-00541-y.


Radiation exposure after myocardial perfusion imaging with Cadmium-Zinc-Telluride camera versus conventional camera.

Baumgarten R, Cerci R, de Nadai Costa A, Pereira-Neto C, Prado P, Zapparoli M J Nucl Cardiol. 2020; 28(3):992-999.

PMID: 32410061 DOI: 10.1007/s12350-020-02146-9.


Update of the Brazilian Guideline on Nuclear Cardiology - 2020.

Mastrocola L, Amorim B, Vitola J, Brandao S, Grossman G, Lima R Arq Bras Cardiol. 2020; 114(2):325-429.

PMID: 32215507 PMC: 7077582. DOI: 10.36660/abc.20200087.


Prospective diagnostic performance of semiconductor SPECT myocardial perfusion imaging: wall thickening analysis reduces the need for an additional prone acquisition.

Djaileb L, Dubois B, De Leiris N, Leenhardt J, Canu M, Sy O Eur J Nucl Med Mol Imaging. 2019; 46(10):2042-2050.

PMID: 31321484 DOI: 10.1007/s00259-019-04415-3.


References
1.
Ben-Haim S, Murthy V, Breault C, Allie R, Sitek A, Roth N . Quantification of Myocardial Perfusion Reserve Using Dynamic SPECT Imaging in Humans: A Feasibility Study. J Nucl Med. 2013; 54(6):873-9. PMC: 3951831. DOI: 10.2967/jnumed.112.109652. View

2.
Esteves F, Raggi P, Folks R, Keidar Z, Askew J, Rispler S . Novel solid-state-detector dedicated cardiac camera for fast myocardial perfusion imaging: multicenter comparison with standard dual detector cameras. J Nucl Cardiol. 2009; 16(6):927-34. PMC: 2776146. DOI: 10.1007/s12350-009-9137-2. View

3.
Duvall W, Croft L, Godiwala T, Ginsberg E, George T, Henzlova M . Reduced isotope dose with rapid SPECT MPI imaging: initial experience with a CZT SPECT camera. J Nucl Cardiol. 2010; 17(6):1009-14. DOI: 10.1007/s12350-010-9215-5. View

4.
Gheorghiade M, Bonow R . Chronic heart failure in the United States: a manifestation of coronary artery disease. Circulation. 1998; 97(3):282-9. DOI: 10.1161/01.cir.97.3.282. View

5.
Rajaram M, Tahari A, Lee A, Lodge M, Tsui B, Nekolla S . Cardiac PET/CT misregistration causes significant changes in estimated myocardial blood flow. J Nucl Med. 2012; 54(1):50-4. DOI: 10.2967/jnumed.112.108183. View