» Articles » PMID: 17506882

Experimental and Analytical Validation of a Modular Acetabular Prosthesis in Total Hip Arthroplasty

Overview
Publisher Biomed Central
Specialty Orthopedics
Date 2007 May 18
PMID 17506882
Citations 3
Authors
Affiliations
Soon will be listed here.
Abstract

A finite element model has been developed to predict in vivo micro motion between a modular acetabular cup and liner after cement less total hip arthroplasty. The purpose of this study is to experimentally validate the model. Six LVDT sensors were used to monitor the micromotion of the liner when subjected to loading conditions ranging from 250 N to 5000 N. Deformations at points of interest for both the experiment and FEM were compared. Results of the FEM with different coefficient of friction between the liner and the cup were investigated to correlate with the experimental results.

Citing Articles

Head, acetabular liner composition, and rate of revision and wear in total hip arthroplasty: a Bayesian network meta-analysis.

Merfort R, Maffulli N, Hofmann U, Hildebrand F, Simeone F, Eschweiler J Sci Rep. 2023; 13(1):20327.

PMID: 37989863 PMC: 10663607. DOI: 10.1038/s41598-023-47670-z.


and methods for the biomechanical assessment of osseointegrated transfemoral prostheses: a systematic review.

Galteri G, Cristofolini L Front Bioeng Biotechnol. 2023; 11:1237919.

PMID: 37662439 PMC: 10469938. DOI: 10.3389/fbioe.2023.1237919.


Contact mechanics of modular metal-on-polyethylene total hip replacement under adverse edge loading conditions.

Hua X, Li J, Wang L, Jin Z, Wilcox R, Fisher J J Biomech. 2014; 47(13):3303-9.

PMID: 25218504 PMC: 4199141. DOI: 10.1016/j.jbiomech.2014.08.015.

References
1.
Morlock M, Schneider E, Bluhm A, Vollmer M, Bergmann G, Muller V . Duration and frequency of every day activities in total hip patients. J Biomech. 2001; 34(7):873-81. DOI: 10.1016/s0021-9290(01)00035-5. View

2.
Saikko V, Calonius O . Slide track analysis of the relative motion between femoral head and acetabular cup in walking and in hip simulators. J Biomech. 2002; 35(4):455-64. DOI: 10.1016/s0021-9290(01)00224-x. View

3.
Illgen 2nd R, Rubash H . The optimal fixation of the cementless acetabular component in primary total hip arthroplasty. J Am Acad Orthop Surg. 2002; 10(1):43-56. DOI: 10.5435/00124635-200201000-00007. View

4.
Dowdy P, Rorabeck C, Bourne R . Uncemented total hip arthroplasty in patients 50 years of age or younger. J Arthroplasty. 1998; 12(8):853-62. DOI: 10.1016/s0883-5403(97)90154-6. View

5.
Bergmann G, Deuretzbacher G, Heller M, Graichen F, Rohlmann A, Strauss J . Hip contact forces and gait patterns from routine activities. J Biomech. 2001; 34(7):859-71. DOI: 10.1016/s0021-9290(01)00040-9. View