The Effect of Conformity and Plastic Thickness on Contact Stresses in Metal-backed Plastic Implants

Overview
Journal J Biomech Eng
Date 1985 Aug 1
PMID 4046559
Citations 15
Authors
Affiliations
Soon will be listed here.
Abstract

Surface damage in polyethylene components for total joint replacement is associated with large contact stresses. An elasticity solution and finite element analyses were used to determine the influence of design parameters on the stresses due to contact in metal-backed components. For nearly conforming contact surfaces, it was found that the stresses in the plastic are very sensitive to clearance, that minimum plastic thickness of 4-6 mm should be maintained for metal-backed components, and that bonding the plastic to the metal backing reduces tensile stresses in the plastic at the edge of the contact zone.

Citing Articles

Investigation of contact behavior on a model of the dual-mobility artificial hip joint for Asians in different inner liner thicknesses.

Hidayat T, Ammarullah M, Ismail R, Saputra E, Lamura M, K N C World J Orthop. 2024; 15(4):321-336.

PMID: 38680676 PMC: 11045469. DOI: 10.5312/wjo.v15.i4.321.


Numerical investigation on the wear characteristics of hip implant under static loading.

Abu Qudeiri J, Abdudeen A, Sahadevan M, Padmanabhan M A Heliyon. 2024; 10(4):e26151.

PMID: 38404850 PMC: 10884854. DOI: 10.1016/j.heliyon.2024.e26151.


In vivo creep and wear performance of vitamin-E-diffused highly crosslinked polyethylene in total hip arthroplasty.

Matsuoka T, Takahashi Y, Ishida T, Tateiwa T, Shishido T, Yamamoto K Arch Orthop Trauma Surg. 2023; 143(12):7195-7203.

PMID: 37438580 DOI: 10.1007/s00402-023-04972-1.


Wear Risk Prevention and Reduction in Total Hip Arthroplasty. A Personalized Study Comparing Cement and Cementless Fixation Techniques Employing Finite Element Analysis.

Gonzalez-Bravo C, Ortega M, Bujan J, de la Torre B, Barrios L J Pers Med. 2021; 11(8).

PMID: 34442424 PMC: 8402105. DOI: 10.3390/jpm11080780.


A Combined Geometric Morphometric and Discrete Element Modeling Approach for Hip Cartilage Contact Mechanics.

Van Houcke J, Audenaert E, Atkins P, Anderson A Front Bioeng Biotechnol. 2020; 8:318.

PMID: 32373602 PMC: 7186355. DOI: 10.3389/fbioe.2020.00318.