Background: OCT can image plaque microstructure at a level of resolution not previously demonstrated with other imaging techniques because it uses infrared light rather than acoustic waves.

Objectives: To compare optical coherence tomography (OCT) and intravascular ultrasound (IVUS) imaging of in vitro atherosclerotic plaques.

Methods: Segments of abdominal aorta were obtained immediately before postmortem examination. Images of 20 sites from five patients were acquired with OCT (operating at an optical wavelength of 1300 nm which was delivered to the sample through an optical fibre) and a 30 MHz ultrasonic transducer. After imaging, the microstructure of the tissue was assessed by routine histological processing.

Results: OCT yielded superior structural information in all plaques examined. The mean (SEM) axial resolution of OCT and IVUS imaging was 16 (1) and 110 (7), respectively, as determined by the point spread function from a mirror. Furthermore, the dynamic range of OCT was 109 dB compared with 43 dB for IVUS imaging.

Conclusions: OCT represents a promising new technology for intracoronary imaging because of its high resolution, broad dynamic range, and ability to be delivered through intravascular catheters.

Citing Articles

A Backing-Layer-Shared Miniature Dual-Frequency Ultrasound Probe for Intravascular Ultrasound Imaging: In Vitro and Ex Vivo Validations.

He Y, Liu X, Zhang J, Peng C Biosensors (Basel). 2023; 13(11).

PMID: 37998146 PMC: 10669229. DOI: 10.3390/bios13110971.


Optical coherence tomography in the diagnosis of myocardial infarction with non-obstructive coronary arteries.

Iwanczyk S, Wozniak P, Araszkiewicz A, Grygier M, Klotzka A, Lesiak M Postepy Kardiol Interwencyjnej. 2023; 18(3):192-200.

PMID: 36751279 PMC: 9885232. DOI: 10.5114/aic.2022.121233.


Automated analysis of fibrous cap in intravascular optical coherence tomography images of coronary arteries.

Lee J, Pereira G, Gharaibeh Y, Kolluru C, Zimin V, Dallan L Sci Rep. 2022; 12(1):21454.

PMID: 36509806 PMC: 9744742. DOI: 10.1038/s41598-022-24884-1.


Vulnerable Plaque, Characteristics, Detection, and Potential Therapies.

Hafiane A J Cardiovasc Dev Dis. 2019; 6(3).

PMID: 31357630 PMC: 6787609. DOI: 10.3390/jcdd6030026.


Development of High-Frequency (>60 MHz) Intravascular Ultrasound (IVUS) Transducer by Using Asymmetric Electrodes for Improved Beam Profile.

Sung J, Jeong J Sensors (Basel). 2018; 18(12).

PMID: 30551639 PMC: 6308511. DOI: 10.3390/s18124414.


References
1.
Peters R, Kok W, Havenith M, Rijsterborgh H, van der Wal A, Visser C . Histopathologic validation of intracoronary ultrasound imaging. J Am Soc Echocardiogr. 1994; 7(3 Pt 1):230-41. DOI: 10.1016/s0894-7317(14)80393-3. View

2.
Finet G, Maurincomme E, Tabib A, Crowley R, Magnin I, Roriz R . Artifacts in intravascular ultrasound imaging: analyses and implications. Ultrasound Med Biol. 1993; 19(7):533-47. DOI: 10.1016/0301-5629(93)90077-2. View

3.
Puliafito C, Hee M, Lin C, Reichel E, Schuman J, Duker J . Imaging of macular diseases with optical coherence tomography. Ophthalmology. 1995; 102(2):217-29. DOI: 10.1016/s0161-6420(95)31032-9. View

4.
Hee M, Izatt J, Swanson E, Huang D, Schuman J, Lin C . Optical coherence tomography of the human retina. Arch Ophthalmol. 1995; 113(3):325-32. DOI: 10.1001/archopht.1995.01100030081025. View

5.
Lee D, Eigler N, Luo H, Nishioka T, Tabak S, Forrester J . Effect of intracoronary ultrasound imaging on clinical decision making. Am Heart J. 1995; 129(6):1084-93. DOI: 10.1016/0002-8703(95)90387-9. View