» Articles » PMID: 38345699

Association of Refractive Outcome with Postoperative Anterior Chamber Depth Measured with 3 Optical Biometers

Overview
Journal Int Ophthalmol
Specialty Ophthalmology
Date 2024 Feb 12
PMID 38345699
Authors
Affiliations
Soon will be listed here.
Abstract

Purpose: This study evaluated the relationship between refractive outcomes and postoperative anterior chamber depth (ACD, measured from corneal epithelium to lens) measured by swept-source optical coherence tomography (SS-OCT), optical low-coherence reflectometry (OLCR), and Scheimpflug devices under the undilated pupil.

Methods: Patients undergoing cataract phacoemulsification with intraocular lens (IOL) implantation in a hospital setting were enrolled. Postoperative ACD (postACD) was performed with an SS-OCT device, an OLCR device, and a Scheimpflug device at least 1 month after cataract surgery. After adjusting the mean predicted error to 0, differences in refractive outcomes were calculated with the Olsen formula using actual postACD measured from 3 devices and predicted value.

Results: Overall, this comparative case study included 69 eyes of 69 patients, and postACD measurements were successfully taken using all 3 devices. The postACD measured with the SS-OCT, OLCR, and Scheimpflug devices was 4.59 ± 0.30, 4.50 ± 0.30, and 4.54 ± 0.32 mm, respectively. Statistically significant differences in postACD were found among 3 devices (P < 0.001), with intraclass correlation coefficients (ICCs) and Bland-Altman showing good agreement. No significant difference in median absolute error was found with the Olsen formula using actual postACD obtained with 3 devices. Percentage prediction errors were within ± 0.50 D in 65% (OLCR), 70% (Scheimpflug), and 67% (SS-OCT) calculated by actual postACD versus 64% by predicted value.

Conclusion: Substantial agreement was found in postACD measurements obtained from the SS-OCT, OLCR, and Scheimpflug devices, with a trend toward comparable refractive outcomes in the Olsen formula. Meanwhile, postACD measurements may be potentially superior for the additional enhancement of refractive outcomes.

References
1.
Olsen T . Sources of error in intraocular lens power calculation. J Cataract Refract Surg. 1992; 18(2):125-9. DOI: 10.1016/s0886-3350(13)80917-0. View

2.
Norrby S . Sources of error in intraocular lens power calculation. J Cataract Refract Surg. 2008; 34(3):368-76. DOI: 10.1016/j.jcrs.2007.10.031. View

3.
Savini G, Hoffer K, Carbonelli M . Anterior chamber and aqueous depth measurement in pseudophakic eyes: agreement between ultrasound biometry and scheimpflug imaging. J Refract Surg. 2013; 29(2):121-5. DOI: 10.3928/1081597X-20130117-07. View

4.
Olsen T . Calculation of intraocular lens power: a review. Acta Ophthalmol Scand. 2007; 85(5):472-85. DOI: 10.1111/j.1600-0420.2007.00879.x. View

5.
Olsen T . Prediction of the effective postoperative (intraocular lens) anterior chamber depth. J Cataract Refract Surg. 2006; 32(3):419-24. DOI: 10.1016/j.jcrs.2005.12.139. View