High Precision Biometry of Pseudophakic Eyes Using Partial Coherence Interferometry
Overview
Affiliations
Purpose: To investigate the applicability of the scanning version of dual-beam partial coherence interferometry (PCI) for measuring the anterior segment and axial length of pseudophakic eyes in a clinical setting and to determine the achievable precision with this biometry technique.
Setting: Department of Ophthalmology, Vienna General Hospital, and Institute of Medical Physics, University of Vienna, Austria.
Methods: Partial coherence interferometry was performed in 39 pseudophakic eyes of 39 patients after implantation of a foldable acrylic intraocular lens (IOL).
Results: Effective lens position (ELP), IOL thickness and lens-capsule distance (LCD) were determined with a precision of 2 to 3 microns; corneal thickness and axial eye length, with a precision of 0.8 and 5.0 microns, respectively. The mean ELP of the IOL was 4.093 mm +/- 0.290 (SD). In 7 eyes (18%), a positive LCD of 68 +/- 40 microns was detected with PCI. Mean corneal thickness was 526.4 +/- 31.5 microns; mean IOL thickness, 791.5 +/- 40.2 microns; and mean axial length, 23.388 +/- 0.824 mm.
Conclusion: The scanning version of PCI enables high precision (< or = 5 microns) and high resolution (approximately 12 microns) biometry of pseudophakic eyes that is better than conventional ultrasound by a factor of more than 20. For the first time, positive LCD, a possible risk factor for posterior capsule opacification, could be detected and quantified. Furthermore, this technique offers a high degree of comfort for the patient since it is a noncontact method with no need for local anesthesia or pupil dilation and has a reduced risk of corneal infection.
Annopawong K, Sriyuttagrai W, Chainakul M, Arjkongharn N, Seresirikachorn K, Kongsomboon K Sci Rep. 2024; 14(1):31026.
PMID: 39730792 PMC: 11681104. DOI: 10.1038/s41598-024-82246-5.
Mao Y, Li J, Qin Y, Xu Y, Liu L, Cheng H Int Ophthalmol. 2024; 44(1):62.
PMID: 38345699 DOI: 10.1007/s10792-024-02995-0.
Optimization of biometry for best refractive outcome in cataract surgery.
Gupta V, Pal H, Sawhney S, Aggarwal A, Vanathi M, Luthra G Indian J Ophthalmol. 2023; 72(1):29-43.
PMID: 38131567 PMC: 10841781. DOI: 10.4103/IJO.IJO_1219_23.
Xiong Y, Lin Y, Zhao Z, Wang H, Zhang G BMC Ophthalmol. 2023; 23(1):218.
PMID: 37194016 PMC: 10189968. DOI: 10.1186/s12886-023-02962-w.
Accuracy of different lens power calculation formulas in patients with phacomorphic glaucoma.
Tabatabaei S, Samadi M, Soleimani M, Fonoodi H, Ghods S, Inanloo B Taiwan J Ophthalmol. 2022; 12(2):164-169.
PMID: 35813804 PMC: 9262030. DOI: 10.4103/tjo.tjo_1_22.