Leaf-individual Calibration for a Double Stack Multileaf Collimator in Photon Radiotherapy

Overview
Specialty Oncology
Date 2023 Aug 28
PMID 37635846
Authors
Affiliations
Soon will be listed here.
Abstract

Background And Purpose: In online adaptive stereotactic body radiotherapy treatments, linear accelerator delivery accuracy is essential. Recently introduced double stack multileaf collimators (MLCs) have new facets in their calibration. We established a radiation-based leaf-individual calibration (LIMCA) method for double stack MLCs.

Materials And Methods: MLC leaf positions were evaluated from four cardinal angles with test patterns at measurement positions throughout the radiation field on EBT3 radiochromic film for each single stack. The accuracy of the method and repeatability of the results were assessed. The effect of MLC positioning errors was characterized for a measured output factor curve and a clinical patient plan.

Results: All positions in the motor step - position calibration file were optimized in the established LIMCA method. The resulting double stack mean accuracy for all angles was 0.2 ± 0.1 mm for X1 (left bank) and 0.2 ± 0.2 mm for X2 (right bank). The accuracy of the leaf position evaluation was 0.2 mm (95% confidence level). The MLC calibration remained stable over four months. Small MLC leaf position errors (e.g. 1.2 mm field size reduction) resulted in important dose errors (-5.8 %) for small quadratic fields of 0.83 × 0.83 cm. Single stack position accuracy was essential for highly modulated treatment plans.

Conclusions: LIMCA is a new double stack MLC calibration method that increases treatment accuracy from four angles and for all moving leaves.

References
1.
Fiandra C, Ricardi U, Ragona R, Anglesio S, Romana Giglioli F, Calamia E . Clinical use of EBT model Gafchromic film in radiotherapy. Med Phys. 2006; 33(11):4314-9. DOI: 10.1118/1.2362876. View

2.
Hoegen P, Zhang K, Tonndorf-Martini E, Weykamp F, Regnery S, Naumann P . MR-guided adaptive versus ITV-based stereotactic body radiotherapy for hepatic metastases (MAESTRO): a randomized controlled phase II trial. Radiat Oncol. 2022; 17(1):59. PMC: 8958771. DOI: 10.1186/s13014-022-02033-2. View

3.
Zwan B, Barnes M, Fuangord T, Stanton C, OConnor D, Keall P . An EPID-based system for gantry-resolved MLC quality assurance for VMAT. J Appl Clin Med Phys. 2016; 17(5):348-365. PMC: 5874117. DOI: 10.1120/jacmp.v17i5.6312. View

4.
Sanchez-Doblado F, Hartmann G, Pena J, Capote R, Paiusco M, Rhein B . Uncertainty estimation in intensity-modulated radiotherapy absolute dosimetry verification. Int J Radiat Oncol Biol Phys. 2007; 68(1):301-10. DOI: 10.1016/j.ijrobp.2006.11.056. View

5.
Rippke C, Schrenk O, Renkamp C, Buchele C, Horner-Rieber J, Debus J . Quality assurance for on-table adaptive magnetic resonance guided radiation therapy: A software tool to complement secondary dose calculation and failure modes discovered in clinical routine. J Appl Clin Med Phys. 2022; 23(3):e13523. PMC: 8906229. DOI: 10.1002/acm2.13523. View