» Articles » PMID: 37536665

Validation of Wearable Digital Devices for Heart Rate Measurement During Exercise Test in Patients With Coronary Artery Disease

Overview
Journal Ann Rehabil Med
Date 2023 Aug 3
PMID 37536665
Authors
Affiliations
Soon will be listed here.
Abstract

Objective: To assess the accuracy of recently commercialized wearable devices in heart rate (HR) measurement during cardiopulmonary exercise test (CPX) under gradual increase in exercise intensity, while wearable devices with HR monitors are reported to be less accurate in different exercise intensities.

Methods: CPX was performed for patients with coronary artery disease (CAD). Twelve lead electrocardiograph (ECG) was the gold standard and Apple watch 7 (AW7), Galaxy watch 4 (GW4) and Bio Patch Mobicare 200 (MC200) were applied for comparison. Paired absolute difference (PAD), mean absolute percentage error (MAPE) and intraclass correlation coefficient (ICC) were evaluated for each device.

Results: Forty-four participants with CAD were included. All the devices showed MAPE under 2% and ICC above 0.9 in rest, exercise and recovery phases (MC200=0.999, GW4=0.997, AW7=0.998). When comparing exercise and recovery phase, PAD of MC200 and AW7 in recovery phase were significantly bigger than PAD of exercise phase (p<0.05). Although not significant, PAD of GW4 tended to be bigger in recovery phase, too. Also, when stratified by HR 20, ICC of all the devices were highest under HR of 100, and ICC decreased as HR increased. However, except for ICC of GW4 at HR above 160 (=0.867), all ICCs exceeded 0.9 indicating excellent accuracy.

Conclusion: The HR measurement of the devices validated in this study shows a high concordance with the ECG device, so CAD patients may benefit from the devices during high-intensity exercise under conditions where HR is measured reliably.

Citing Articles

Prognostic Factors for Responders of Home-Based Pulmonary Rehabilitation-Secondary Analysis of a Randomized Controlled Trial.

Kim C, Choi H, Rhee C, Lee J, Oh J, Song J Healthcare (Basel). 2025; 13(3).

PMID: 39942497 PMC: 11817295. DOI: 10.3390/healthcare13030308.


The Accessibility and Effect of Cardiac Rehabilitation in COVID-19 Pandemic Era.

Kim C, Song J, Kim S Ann Rehabil Med. 2024; 48(4):249-258.

PMID: 39074836 PMC: 11372283. DOI: 10.5535/arm.240021.

References
1.
Antoniou V, Davos C, Kapreli E, Batalik L, Panagiotakos D, Pepera G . Effectiveness of Home-Based Cardiac Rehabilitation, Using Wearable Sensors, as a Multicomponent, Cutting-Edge Intervention: A Systematic Review and Meta-Analysis. J Clin Med. 2022; 11(13). PMC: 9267864. DOI: 10.3390/jcm11133772. View

2.
Etiwy M, Akhrass Z, Gillinov L, Alashi A, Wang R, Blackburn G . Accuracy of wearable heart rate monitors in cardiac rehabilitation. Cardiovasc Diagn Ther. 2019; 9(3):262-271. PMC: 6603497. DOI: 10.21037/cdt.2019.04.08. View

3.
Koo T, Li M . A Guideline of Selecting and Reporting Intraclass Correlation Coefficients for Reliability Research. J Chiropr Med. 2016; 15(2):155-63. PMC: 4913118. DOI: 10.1016/j.jcm.2016.02.012. View

4.
Bots S, Peters S, Woodward M . Sex differences in coronary heart disease and stroke mortality: a global assessment of the effect of ageing between 1980 and 2010. BMJ Glob Health. 2017; 2(2):e000298. PMC: 5435266. DOI: 10.1136/bmjgh-2017-000298. View

5.
Alzahrani A, Hu S, Azorin-Peris V, Barrett L, Esliger D, Hayes M . A multi-channel opto-electronic sensor to accurately monitor heart rate against motion artefact during exercise. Sensors (Basel). 2015; 15(10):25681-702. PMC: 4634442. DOI: 10.3390/s151025681. View