Accuracy of Wearable Heart Rate Monitors in Cardiac Rehabilitation

Abstract

Background: To assess the accuracy of four wearable heart rate (HR) monitors in patients with established cardiovascular disease enrolled in phase II or III cardiac rehabilitation (CR).

Methods: Eighty adult patients enrolled in phase II or III CR were monitored during a CR session that included exercise on a treadmill and/or stationary cycle. Participants underwent HR monitoring with standard ECG limb leads, an electrocardiographic (ECG) chest strap monitor (Polar H7), and two randomly assigned wrist-worn HR monitors (Apple Watch, Fitbit Blaze, Garmin Forerunner 235, TomTom Spark Cardio), one on each wrist. HR was recorded at rest and at 3, 5, and 7 minutes of steady-state exercise on the treadmill and stationary cycle.

Results: Across all exercise conditions, the chest strap monitor (Polar H7) had the best agreement with ECG (r=0.99) followed by the Apple Watch (r=0.80), Fitbit Blaze (r=0.78), TomTom Spark (r=0.76) and Garmin Forerunner (r=0.52). There was variability in accuracy under different exercise conditions. On the treadmill, only the Fitbit Blaze performed well (r=0.76), while on the stationary cycle, Apple Watch (r=0.89) and TomTom Spark (r=0.85) were most accurate.

Conclusions: In cardiac patients, the accuracy of wearable, optically based HR monitors varies, and none of those tested was as accurate as an electrode-containing chest monitor. This observation has implications for in-home CR, as electrode-containing chest monitors should be used when accurate HR measurement is imperative.

Citing Articles

Wearable devices for patient monitoring in the intensive care unit.

Angelucci A, Greco M, Cecconi M, Aliverti A Intensive Care Med Exp. 2025; 13(1):26.

PMID: 40016479 PMC: 11868008. DOI: 10.1186/s40635-025-00738-8.


Validity and Inter-Device Reliability of the OTBeat Burn Monitor to Estimate Heart Rate During Exercise.

de Leon M, Camic C, Herring M, Hill C Sports (Basel). 2025; 13(2).

PMID: 39997980 PMC: 11860634. DOI: 10.3390/sports13020049.


Validity of heart rate measurements in wrist-based monitors across skin tones during exercise.

Hung S, Serwa K, Rosenthal G, Eng J PLoS One. 2025; 20(2):e0318724.

PMID: 39928630 PMC: 11809914. DOI: 10.1371/journal.pone.0318724.


Accuracy of Apple Watch in Measuring 30-Second Resting Electrocardiography in Patients with Cardiac Diseases and Comorbidity: An Observational Cross-Sectional Study.

Khushhal A, Mohamed A, Alsegame M, Alsaedi A J Multidiscip Healthc. 2025; 18:493-504.

PMID: 39906073 PMC: 11790806. DOI: 10.2147/JMDH.S507504.


Efficacy of Wearable Single-Lead ECG Monitoring during Exercise Stress Testing: A Comparative Study.

Choi H, Lee S, Choi J, Kim G, Lee Y, Lee J Sensors (Basel). 2024; 24(19).

PMID: 39409434 PMC: 11479017. DOI: 10.3390/s24196394.


References
1.
OConnor C, Whellan D, Lee K, Keteyian S, Cooper L, Ellis S . Efficacy and safety of exercise training in patients with chronic heart failure: HF-ACTION randomized controlled trial. JAMA. 2009; 301(14):1439-50. PMC: 2916661. DOI: 10.1001/jama.2009.454. View

2.
Lawler P, Filion K, Eisenberg M . Efficacy of exercise-based cardiac rehabilitation post-myocardial infarction: a systematic review and meta-analysis of randomized controlled trials. Am Heart J. 2011; 162(4):571-584.e2. DOI: 10.1016/j.ahj.2011.07.017. View

3.
Smith Jr S, Benjamin E, Bonow R, Braun L, Creager M, Franklin B . AHA/ACCF secondary prevention and risk reduction therapy for patients with coronary and other atherosclerotic vascular disease: 2011 update: a guideline from the American Heart Association and American College of Cardiology Foundation endorsed by.... J Am Coll Cardiol. 2011; 58(23):2432-46. DOI: 10.1016/j.jacc.2011.10.824. View

4.
Yancy C, Jessup M, Bozkurt B, Butler J, Casey Jr D, Drazner M . 2013 ACCF/AHA guideline for the management of heart failure: a report of the American College of Cardiology Foundation/American Heart Association Task Force on Practice Guidelines. J Am Coll Cardiol. 2013; 62(16):e147-239. DOI: 10.1016/j.jacc.2013.05.019. View

5.
Clark R, Conway A, Poulsen V, Keech W, Tirimacco R, Tideman P . Alternative models of cardiac rehabilitation: a systematic review. Eur J Prev Cardiol. 2013; 22(1):35-74. DOI: 10.1177/2047487313501093. View