» Articles » PMID: 36198027

Accelerated 4D-flow MRI with 3-point Encoding Enabled By machine Learning

Overview
Journal Magn Reson Med
Publisher Wiley
Specialty Radiology
Date 2022 Oct 5
PMID 36198027
Authors
Affiliations
Soon will be listed here.
Abstract

Purpose: To investigate the acceleration of 4D-flow MRI using a convolutional neural network (CNN) that produces three directional velocities from three flow encodings, without requiring a fourth reference scan measuring background phase.

Methods: A fully 3D CNN using a U-net architecture was trained in a block-wise fashion to take complex images from three flow encodings and to produce three real-valued images for each velocity component. Using neurovascular 4D-flow scans (n = 144), the CNN was trained to predict velocities computed from four flow encodings by standard reconstruction including correction for residual background phase offsets. Methods to optimize loss functions were investigated, including magnitude, complex difference, and uniform velocity weightings. Subsequently, 3-point encoding was evaluated using cross validation of pixelwise correlation, flow measurements in major arteries, and in experiments with data at differing acceleration rates than the training data.

Results: The CNN-produced 3-point velocities showed excellent agreements with the 4-point velocities, both qualitatively in velocity images, in flow rate measures, and quantitatively in regression analysis (slope = 0.96, R  = 0.992). Optimizing the training to focus on vessel velocities rather than the global velocity error and improved the correlation of velocity within vessels themselves. The lowest error was observed when the loss function used uniform velocity weighting, in which the magnitude-weighted inverse of the velocity frequency uniformly distributed weighting across all velocity ranges. When applied to highly accelerated data, the 3-point network maintained a high correlation with ground truth data and demonstrated a denoising effect.

Conclusion: The 4D-flow MRI can be accelerated using machine learning requiring only three flow encodings to produce three-directional velocity maps with small errors.

Citing Articles

Macrovascular blood flow and microvascular cerebrovascular reactivity are regionally coupled in adolescence.

Zvolanek K, Moore J, Jarvis K, Moum S, Bright M J Cereb Blood Flow Metab. 2024; :271678X241298588.

PMID: 39534950 PMC: 11563552. DOI: 10.1177/0271678X241298588.


A review of deep learning-based reconstruction methods for accelerated MRI using spatiotemporal and multi-contrast redundancies.

Kim S, Park H, Park S Biomed Eng Lett. 2024; 14(6):1221-1242.

PMID: 39465106 PMC: 11502678. DOI: 10.1007/s13534-024-00425-9.


Macrovascular blood flow and microvascular cerebrovascular reactivity are regionally coupled in adolescence.

Zvolanek K, Moore J, Jarvis K, Moum S, Bright M bioRxiv. 2024; .

PMID: 38746187 PMC: 11092525. DOI: 10.1101/2024.04.26.590312.


Four-dimensional flow MRI for quantitative assessment of cerebrospinal fluid dynamics: Status and opportunities.

Rivera-Rivera L, Vikner T, Eisenmenger L, Johnson S, Johnson K NMR Biomed. 2023; 37(7):e5082.

PMID: 38124351 PMC: 11162953. DOI: 10.1002/nbm.5082.


Deep Learning-Based Reconstruction for Cardiac MRI: A Review.

Oscanoa J, Middione M, Alkan C, Yurt M, Loecher M, Vasanawala S Bioengineering (Basel). 2023; 10(3).

PMID: 36978725 PMC: 10044915. DOI: 10.3390/bioengineering10030334.


References
1.
Kim D, Jen M, Eisenmenger L, Johnson K . Accelerated 4D-flow MRI with 3-point encoding enabled by machine learning. Magn Reson Med. 2022; 89(2):800-811. PMC: 9712238. DOI: 10.1002/mrm.29469. View

2.
Rivera-Rivera L, Cody K, Eisenmenger L, Cary P, Rowley H, Carlsson C . Assessment of vascular stiffness in the internal carotid artery proximal to the carotid canal in Alzheimer's disease using pulse wave velocity from low rank reconstructed 4D flow MRI. J Cereb Blood Flow Metab. 2020; 41(2):298-311. PMC: 8370001. DOI: 10.1177/0271678X20910302. View

3.
Pelc N, Bernstein M, Shimakawa A, Glover G . Encoding strategies for three-direction phase-contrast MR imaging of flow. J Magn Reson Imaging. 1991; 1(4):405-13. DOI: 10.1002/jmri.1880010404. View

4.
Wang D, Chien A, Shao J, Ali F, Hu P . Accelerated phase contrast MRI using hybrid one- and two-sided flow encodings only (HOTFEO). NMR Biomed. 2018; 31(5):e3904. DOI: 10.1002/nbm.3904. View

5.
Ma L, Markl M, Chow K, Huh H, Forman C, Vali A . Aortic 4D flow MRI in 2 minutes using compressed sensing, respiratory controlled adaptive k-space reordering, and inline reconstruction. Magn Reson Med. 2019; 81(6):3675-3690. PMC: 6535305. DOI: 10.1002/mrm.27684. View