Concomitant Gradient Terms in Phase Contrast MR: Analysis and Correction

Overview
Journal Magn Reson Med
Publisher Wiley
Specialty Radiology
Date 1998 Feb 20
PMID 9469714
Citations 208
Authors
Affiliations
Soon will be listed here.
Abstract

Whenever a linear gradient is activated, concomitant magnetic fields with non-linear spatial dependence result. This is a consequence of Maxwell's equations, i.e., within the imaging volume the magnetic field must have zero divergence, and has negligible curl. The concomitant, or Maxwell field has been described in the MRI literature for over 10 years. In this paper, we theoretically and experimentally show the existence of two additional lowest-order terms in the concomitant field, which we call cross-terms. The concomitant gradient cross-terms only arise when the longitudinal gradient Gz is simultaneously active with a transverse gradient (Gx or Gy). The effect of all of the concomitant gradient terms on phase contrast imaging is examined in detail. Several methods for reducing or eliminating phase errors arising from the concomitant magnetic field are described. The feasibility of a joint pulse sequence-reconstruction method, which requires no increase in minimum TE, is demonstrated. Since the lowest-order terms of the concomitant field are proportional to G2/B0, the importance of concomitant gradient terms is expected to increase given the current interest in systems with stronger gradients and/or weaker main magnetic fields.

Citing Articles

Respiratory influence on cerebral blood flow and blood volume - A 4D flow MRI study.

Soderstrom P, Eklund A, Karalija N, Andersson B, Riklund K, Backman L J Cereb Blood Flow Metab. 2025; :271678X251316395.

PMID: 39883800 PMC: 11783418. DOI: 10.1177/0271678X251316395.


Phantom-based gradient waveform measurements with compensated variable-prephasing: Description and application to EPI at 7 T.

Scholten H, Wech T, Homolya I, Kostler H Magn Reson Med. 2025; 93(5):2209-2223.

PMID: 39834125 PMC: 11893026. DOI: 10.1002/mrm.30425.


Robustness of 4D flow MRI derived aortic wall shear stress and pulse wave velocity across different protocols in healthy controls and in patients with bicuspid aortic valve.

Dushfunian D, Maroun A, Berhan H, Baraboo J, Johnson E, Jarvis K Int J Cardiovasc Imaging. 2024; 41(1):137-149.

PMID: 39652207 DOI: 10.1007/s10554-024-03299-1.


3D MR elastography at 0.55 T: Concomitant field effects and feasibility.

Darwish O, Di Cio P, Sinkus R, Neji R Magn Reson Med. 2024; 93(4):1602-1614.

PMID: 39587762 PMC: 11782726. DOI: 10.1002/mrm.30377.


Macrovascular blood flow and microvascular cerebrovascular reactivity are regionally coupled in adolescence.

Zvolanek K, Moore J, Jarvis K, Moum S, Bright M J Cereb Blood Flow Metab. 2024; :271678X241298588.

PMID: 39534950 PMC: 11563552. DOI: 10.1177/0271678X241298588.