Pulse Arrival Time As a Surrogate of Blood Pressure

Overview
Journal Sci Rep
Specialty Science
Date 2021 Nov 24
PMID 34815419
Citations 24
Authors
Affiliations
Soon will be listed here.
Abstract

Various models have been proposed for the estimation of blood pressure (BP) from pulse transit time (PTT). PTT is defined as the time delay of the pressure wave, produced by left ventricular contraction, measured between a proximal and a distal site along the arterial tree. Most researchers, when they measure the time difference between the peak of the R-wave in the electrocardiogram signal (corresponding to left ventricular depolarisation) and a fiducial point in the photoplethysmogram waveform (as measured by a pulse oximeter attached to the fingertip), describe this erroneously as the PTT. In fact, this is the pulse arrival time (PAT), which includes not only PTT, but also the time delay between the electrical depolarisation of the heart's left ventricle and the opening of the aortic valve, known as pre-ejection period (PEP). PEP has been suggested to present a significant limitation to BP estimation using PAT. This work investigates the impact of PEP on PAT, leading to a discussion on the best models for BP estimation using PAT or PTT. We conducted a clinical study involving 30 healthy volunteers (53.3% female, 30.9 ± 9.35 years old, with a body mass index of 22.7 ± 3.2 kg/m[Formula: see text]). Each session lasted on average 27.9 ± 0.6 min and BP was varied by an infusion of phenylephrine (a medication that causes venous and arterial vasoconstriction). We introduced new processing steps for the analysis of PAT and PEP signals. Various population-based models (Poon, Gesche and Fung) and a posteriori models (inverse linear, inverse squared and logarithm) for estimation of BP from PTT or PAT were evaluated. Across the cohort, PEP was found to increase by 5.5 ms ± 4.5 ms from its baseline value. Variations in PTT were significantly larger in amplitude, - 16.8 ms ± 7.5 ms. We suggest, therefore, that for infusions of phenylephrine, the contribution of PEP on PAT can be neglected. All population-based models produced large BP estimation errors, suggesting that they are insufficient for modelling the complex pathways relating changes in PTT or PAT to changes in BP. Although PAT is inversely correlated with systolic blood pressure (SBP), the gradient of this relationship varies significantly from individual to individual, from - 2946 to - 470.64 mmHg/s in our dataset. For the a posteriori inverse squared model, the root mean squared errors (RMSE) for systolic and diastolic blood pressure (DBP) estimation from PAT were 5.49 mmHg and 3.82 mmHg, respectively. The RMSEs for SBP and DBP estimation by PTT were 4.51 mmHg and 3.53 mmHg, respectively. These models take into account individual calibration curves required for accurate blood pressure estimation. The best performing population-based model (Poon) reported error values around double that of the a posteriori inverse squared model, and so the use of population-based models is not justified.

Citing Articles

AI-driven system for non-contact continuous nocturnal blood pressure monitoring using fiber optic ballistocardiography.

Huang Y, Chen L, Li C, Peng J, Hu Q, Sun Y Commun Eng. 2024; 3(1):183.

PMID: 39702581 PMC: 11659553. DOI: 10.1038/s44172-024-00326-w.


Cardiovascular and vasomotor pulsations in the brain and periphery during awake and NREM sleep in a multimodal fMRI study.

Tuunanen J, Helakari H, Huotari N, Vayrynen T, Jarvela M, Kananen J Front Neurosci. 2024; 18:1457732.

PMID: 39440186 PMC: 11493778. DOI: 10.3389/fnins.2024.1457732.


Wearable Blood Pressure Monitoring Devices: Understanding Heterogeneity in Design and Evaluation.

Cheung M, Sabharwal A, Cote G, Veeraraghavan A IEEE Trans Biomed Eng. 2024; 71(12):3569-3592.

PMID: 39106139 PMC: 11799359. DOI: 10.1109/TBME.2024.3434344.


Non-invasive pulse arrival time is associated with cardiac index in pediatric heart transplant patients with normal ejection fraction.

Kwon S, Weinerman B, Nametz D, Megjhani M, Lee I, Habib A Physiol Meas. 2024; 45(7).

PMID: 38986482 PMC: 11262133. DOI: 10.1088/1361-6579/ad61b9.


Wearable Ring-Shaped Biomedical Device for Physiological Monitoring through Finger-Based Acquisition of Electrocardiographic, Photoplethysmographic, and Galvanic Skin Response Signals: Design and Preliminary Measurements.

Volpes G, Valenti S, Genova G, Bara C, Parisi A, Faes L Biosensors (Basel). 2024; 14(4).

PMID: 38667198 PMC: 11048376. DOI: 10.3390/bios14040205.


References
1.
Balmer J, Pretty C, Davidson S, Desaive T, Kamoi S, Pironet A . Pre-ejection period, the reason why the electrocardiogram Q-wave is an unreliable indicator of pulse wave initialization. Physiol Meas. 2018; 39(9):095005. DOI: 10.1088/1361-6579/aada72. View

2.
Jabalameli M, Rouholamin S, Gourtanian F . A comparison of the effects of fentanyl and remifentanil on nausea, vomiting, and pain after cesarean section. Iran J Med Sci. 2013; 36(3):183-7. PMC: 3556767. View

3.
Kjeldsen S . Hypertension and cardiovascular risk: General aspects. Pharmacol Res. 2017; 129:95-99. DOI: 10.1016/j.phrs.2017.11.003. View

4.
Dolan E, Stanton A, Thijs L, Hinedi K, Atkins N, McClory S . Superiority of ambulatory over clinic blood pressure measurement in predicting mortality: the Dublin outcome study. Hypertension. 2005; 46(1):156-61. DOI: 10.1161/01.HYP.0000170138.56903.7a. View

5.
Vincent J . Understanding cardiac output. Crit Care. 2008; 12(4):174. PMC: 2575587. DOI: 10.1186/cc6975. View