An Impedance Pneumography Signal Quality Index: Design, Assessment and Application to Respiratory Rate Monitoring

Overview
Publisher Elsevier
Date 2021 Jun 25
PMID 34168684
Citations 15
Authors
Affiliations
Soon will be listed here.
Abstract

Impedance pneumography (ImP) is widely used for respiratory rate (RR) monitoring. However, ImP-derived RRs can be imprecise. The aim of this study was to develop a signal quality index (SQI) for the ImP signal, and couple it with a RR algorithm, to improve RR monitoring. An SQI was designed which identifies candidate breaths and assesses signal quality using: the variation in detected breath durations, how well peaks and troughs are defined, and the similarity of breath morphologies. The SQI categorises 32 s signal segments as either high or low quality. Its performance was evaluated using two critical care datasets. RRs were estimated from high-quality segments using a RR algorithm, and compared with reference RRs derived from manual annotations. The SQI had a sensitivity of 77.7 %, and specificity of 82.3 %. RRs estimated from segments classified as high quality were accurate and precise, with mean absolute errors of 0.21 and 0.40 breaths per minute (bpm) on the two datasets. Clinical monitor RRs were significantly less precise. The SQI classified 34.9 % of real-world data as high quality. In conclusion, the proposed SQI accurately identifies high-quality segments, and RRs estimated from those segments are precise enough for clinical decision making. This SQI may improve RR monitoring in critical care. Further work should assess it with wearable sensor data.

Citing Articles

Emerging Wearable Acoustic Sensing Technologies.

Liu T, Mao Y, Dou H, Zhang W, Yang J, Wu P Adv Sci (Weinh). 2025; 12(6):e2408653.

PMID: 39749384 PMC: 11809411. DOI: 10.1002/advs.202408653.


Video-Based Respiratory Rate Estimation for Infants in the NICU.

Ahani S, Niknafs N, Lavoie P, Holsti L, Dumont G IEEE J Transl Eng Health Med. 2024; 12:684-696.

PMID: 39559825 PMC: 11573411. DOI: 10.1109/JTEHM.2024.3488523.


Wearable Bioimpedance Monitoring: Viewpoint for Application in Chronic Conditions.

Groenendaal W, Lee S, Van Hoof C JMIR Biomed Eng. 2024; 6(2):e22911.

PMID: 38907374 PMC: 11041432. DOI: 10.2196/22911.


Respiratory rate measurement by pressure variation in the high flow nasal cannula-system in healthy volunteers.

Miechels J, Koning M J Clin Monit Comput. 2024; 38(6):1397-1404.

PMID: 38867018 DOI: 10.1007/s10877-024-01185-8.


A Novel Respiratory Rate Estimation Algorithm from Photoplethysmogram Using Deep Learning Model.

Chin W, Kwan B, Lim W, Tee Y, Darmaraju S, Liu H Diagnostics (Basel). 2024; 14(3).

PMID: 38337800 PMC: 10855057. DOI: 10.3390/diagnostics14030284.


References
1.
Goldberger A, Amaral L, Glass L, Hausdorff J, Ivanov P, Mark R . PhysioBank, PhysioToolkit, and PhysioNet: components of a new research resource for complex physiologic signals. Circulation. 2000; 101(23):E215-20. DOI: 10.1161/01.cir.101.23.e215. View

2.
Mlynczak M, Niewiadomski W, Zylinski M, Cybulski G . Assessment of calibration methods on impedance pneumography accuracy. Biomed Tech (Berl). 2015; 61(6):587-593. DOI: 10.1515/bmt-2015-0125. View

3.
Schafer A, Kratky K . Estimation of breathing rate from respiratory sinus arrhythmia: comparison of various methods. Ann Biomed Eng. 2008; 36(3):476-85. DOI: 10.1007/s10439-007-9428-1. View

4.
Charlton P, Bonnici T, Tarassenko L, Alastruey J, Clifton D, Beale R . Extraction of respiratory signals from the electrocardiogram and photoplethysmogram: technical and physiological determinants. Physiol Meas. 2017; 38(5):669-690. DOI: 10.1088/1361-6579/aa670e. View

5.
Freundlich J, Erickson J . Electrical impedance pneumography for simple nonrestrictive continuous monitoring of respiratory rate, rhythm and tidal volume for surgical patients. Chest. 1974; 65(2):181-4. DOI: 10.1378/chest.65.2.181. View