A Wearable Multimodal Sensing System for Tracking Changes in Pulmonary Fluid Status, Lung Sounds, and Respiratory Markers

Abstract

Heart failure (HF) exacerbations, characterized by pulmonary congestion and breathlessness, require frequent hospitalizations, often resulting in poor outcomes. Current methods for tracking lung fluid and respiratory distress are unable to produce continuous, holistic measures of cardiopulmonary health. We present a multimodal sensing system that captures bioimpedance spectroscopy (BIS), multi-channel lung sounds from four contact microphones, multi-frequency impedance pneumography (IP), temperature, and kinematics to track changes in cardiopulmonary status. We first validated the system on healthy subjects ( = 10) and then conducted a feasibility study on patients ( = 14) with HF in clinical settings. Three measurements were taken throughout the course of hospitalization, and parameters relevant to lung fluid status-the ratio of the resistances at 5 kHz to those at 150 kHz ()-and respiratory timings (e.g., respiratory rate) were extracted. We found a statistically significant increase in ( < 0.05) from admission to discharge and observed respiratory timings in physiologically plausible ranges. The IP-derived respiratory signals and lung sounds were sensitive enough to detect abnormal respiratory patterns (Cheyne-Stokes) and inspiratory crackles from patient recordings, respectively. We demonstrated that the proposed system is suitable for detecting changes in pulmonary fluid status and capturing high-quality respiratory signals and lung sounds in a clinical setting.

Citing Articles

Digital twin driven electrode optimization for wearable bladder monitoring via bioimpedance.

Crane H, Berkebile J, Mabrouk S, Riccardelli N, Inan O NPJ Digit Med. 2025; 8(1):73.

PMID: 39885231 PMC: 11782588. DOI: 10.1038/s41746-025-01441-4.


Recent Advances in the Wearable Devices for Monitoring and Management of Heart Failure.

Odeh V, Chen Y, Wang W, Ding X Rev Cardiovasc Med. 2024; 25(10):386.

PMID: 39484130 PMC: 11522764. DOI: 10.31083/j.rcm2510386.


A scoping review on advancements in noninvasive wearable technology for heart failure management.

Scholte N, van Ravensberg A, Shakoor A, Boersma E, Ronner E, de Boer R NPJ Digit Med. 2024; 7(1):279.

PMID: 39396094 PMC: 11470936. DOI: 10.1038/s41746-024-01268-5.


Wearable Devices Based on Bioimpedance Test in Heart Failure: Clinical Relevance: Systematic Review.

Gimenez-Miranda L, Scagliusi S, Perez-Garcia P, Olmo-Fernandez A, Huertas G, Yufera A Rev Cardiovasc Med. 2024; 25(9):315.

PMID: 39355607 PMC: 11440438. DOI: 10.31083/j.rcm2509315.


Regional Pulmonary Ventilation Assessment Method and System Based on Impedance Sensing Information from the Pentapulmonary Lobes.

Zhang Y, Song C, He W, Zhang Q, Zhao P, Wang J Sensors (Basel). 2024; 24(10).

PMID: 38794056 PMC: 11124947. DOI: 10.3390/s24103202.


References
1.
Beck C, Georgiou J . Wearable, multimodal, vitals acquisition unit for intelligent field triage. Healthc Technol Lett. 2016; 3(3):189-196. PMC: 5048342. DOI: 10.1049/htl.2016.0038. View

2.
Mukherjee H, Sreerama P, Dhar A, Obaidullah S, Roy K, Mahmud M . Automatic Lung Health Screening Using Respiratory Sounds. J Med Syst. 2021; 45(2):19. PMC: 7797201. DOI: 10.1007/s10916-020-01681-9. View

3.
Klum M, Urban M, Tigges T, Pielmus A, Feldheiser A, Schmitt T . Wearable Cardiorespiratory Monitoring Employing a Multimodal Digital Patch Stethoscope: Estimation of ECG, PEP, LVETand Respiration Using a 55 mm Single-Lead ECG and Phonocardiogram. Sensors (Basel). 2020; 20(7). PMC: 7180963. DOI: 10.3390/s20072033. View

4.
Rolfe S . The importance of respiratory rate monitoring. Br J Nurs. 2019; 28(8):504-508. DOI: 10.12968/bjon.2019.28.8.504. View

5.
Beckmann L, van Riesen D, Leonhardt S . Optimal electrode placement and frequency range selection for the detection of lung water using bioimpedance spectroscopy. Annu Int Conf IEEE Eng Med Biol Soc. 2007; 2007:2685-8. DOI: 10.1109/IEMBS.2007.4352882. View