Comparison of the FreeStyle Libre Pro Flash Continuous Glucose Monitoring (CGM) System and Point-of-Care Capillary Glucose Testing in Hospitalized Patients With Type 2 Diabetes Treated With Basal-Bolus Insulin Regimen

Overview
Journal Diabetes Care
Specialty Endocrinology
Date 2020 Jul 10
PMID 32641372
Citations 72
Authors
Affiliations
Soon will be listed here.
Abstract

Objective: We compared the performance of the FreeStyle Libre Pro continuous glucose monitoring (CGM) and point-of-care capillary glucose testing (POC) among insulin-treated hospitalized patients with type 2 diabetes (T2D).

Research Design And Methods: This was a prospective study in adult patients with T2D admitted to general medicine and surgery wards. Patients were monitored with POC before meals and bedtime and with CGM during the hospital stay. Study end points included differences between POC and CGM in mean daily blood glucose (BG), hypoglycemia <70 and <54 mg/dL, and nocturnal hypoglycemia. We also calculated the mean absolute relative difference (MARD), ±15%/15 mg/dL, ±20%/20 mg/dL, and ±30%/30 mg/dL and error grid analysis between matched glucose pairs.

Results: Mean daily glucose was significantly higher by POC (188.9 ± 37.3 vs. 176.1 ± 46.9 mg/dL) with an estimated mean difference of 12.8 mg/dL (95% CI 8.3-17.2 mg/dL), and proportions of patients with glucose readings <70 mg/dL (14% vs. 56%) and <54 mg/dL (4.1% vs. 36%) detected by POC BG were significantly lower compared with CGM (all < 0.001). Nocturnal and prolonged CGM hypoglycemia <54 mg/dL were 26% and 12%, respectively. The overall MARD was 14.8%, ranging between 11.4% and 16.7% for glucose values between 70 and 250 mg/dL and higher for 51-69 mg/dL (MARD 28.0%). The percentages of glucose readings within ±15%/15 mg/dL, ±20%/20 mg/dL, and ±30%/30 mg/dL were 62%, 76%, and 91%, respectively. Error grid analysis showed 98.8% of glucose pairs within zones A and B.

Conclusions: Compared with POC, FreeStyle Libre CGM showed lower mean daily glucose and higher detection of hypoglycemic events, particularly nocturnal and prolonged hypoglycemia in hospitalized patients with T2D. CGM's accuracy was lower in the hypoglycemic range.

Citing Articles

Continuous Glucose Monitoring in Hospitalized Adults With Diabetic Ketoacidosis: A Prospective Open-Label Pilot Study.

Bogun M, Wang C, Kurlansky P, Bedeir N, Umpierrez G J Diabetes Sci Technol. 2025; :19322968251316887.

PMID: 39907056 PMC: 11800229. DOI: 10.1177/19322968251316887.


Effect of Imeglimin, a Novel Anti-Diabetic Agent, on Insulin Secretion and Glycemic Variability in Type 2 Diabetes Treated with DPP-4 Inhibitor: A 16-Week, Open Label, Pilot Study.

Itsukaichi A, Yoshikawa F, Fuchigami A, Iwata Y, Sato G, Miyagi M Diabetes Metab Syndr Obes. 2025; 18():101-111.

PMID: 39807125 PMC: 11727693. DOI: 10.2147/DMSO.S495930.


Time Below Range and Its Influence on Hypoglycemia Awareness and Severe Hypoglycemia: Insights From the Association of British Clinical Diabetologists Study.

Deshmukh H, Wilmot E, Choudhary P, Ssemmondo E, Barnes D, Walker N Diabetes Care. 2025; 48(3):437-443.

PMID: 39746160 PMC: 11870288. DOI: 10.2337/dc24-1833.


16. Diabetes Care in the Hospital: Standards of Care in Diabetes-2025.

Diabetes Care. 2024; 48(Supplement_1):S321-S334.

PMID: 39651972 PMC: 11635037. DOI: 10.2337/dc25-S016.


A free-living, walking-based, exercise programme, with exercise timed relative to breakfast, to improve metabolic health in people living with overweight and obesity: A feasibility study.

Barrett J, Crozier A, Cuthbertson D, Strauss J, Wagenmakers A, Shepherd S PLoS One. 2024; 19(11):e0307582.

PMID: 39570874 PMC: 11581328. DOI: 10.1371/journal.pone.0307582.


References
1.
Moghissi E, Korytkowski M, DiNardo M, Einhorn D, Hellman R, Hirsch I . American Association of Clinical Endocrinologists and American Diabetes Association consensus statement on inpatient glycemic control. Diabetes Care. 2009; 32(6):1119-31. PMC: 2681039. DOI: 10.2337/dc09-9029. View

2.
Pleus S, Heinemann L, Freckmann G . Blood Glucose Monitoring Data Should Be Reported in Detail When Studies About Efficacy of Continuous Glucose Monitoring Systems Are Published. J Diabetes Sci Technol. 2018; 12(5):1061-1063. PMC: 6134630. DOI: 10.1177/1932296817753629. View

3.
Pasquel F, Spiegelman R, McCauley M, Smiley D, Umpierrez D, Johnson R . Hyperglycemia during total parenteral nutrition: an important marker of poor outcome and mortality in hospitalized patients. Diabetes Care. 2009; 33(4):739-41. PMC: 2845017. DOI: 10.2337/dc09-1748. View

4.
Wallia A, Umpierrez G, Rushakoff R, Klonoff D, Rubin D, Golden S . Consensus Statement on Inpatient Use of Continuous Glucose Monitoring. J Diabetes Sci Technol. 2017; 11(5):1036-1044. PMC: 5950996. DOI: 10.1177/1932296817706151. View

5.
Schaupp L, Donsa K, Neubauer K, Mader J, Aberer F, Holl B . Taking a Closer Look--Continuous Glucose Monitoring in Non-Critically Ill Hospitalized Patients with Type 2 Diabetes Mellitus Under Basal-Bolus Insulin Therapy. Diabetes Technol Ther. 2015; 17(9):611-8. DOI: 10.1089/dia.2014.0343. View