Accuracy of a Real-Time Continuous Glucose Monitor in Pediatric Diabetic Ketoacidosis Admissions

Overview
Date 2024 Mar 5
PMID 38441904
Authors
Affiliations
Soon will be listed here.
Abstract

Continuous glucose monitoring (CGM) devices are integral in the outpatient care of people with type 1 diabetes, although they lack inpatient labeling. Food and Drug Administration began allowing inpatient use during the coronavirus disease 2019 (COVID-19) pandemic, with some accuracy data now available, primarily from adult hospitals. Pediatric inpatient data remain limited, particularly during diabetic ketoacidosis (DKA) admissions and for patients receiving intravenous (IV) insulin. This retrospective chart review compared point-of-care glucose values to personal Dexcom G6 sensor data during pediatric hospitalizations. Accuracy was assessed using mean absolute relative difference (MARD), Clarke Error Grids, and the percentage of values within 15/20/30% if glucose value >100 mg/dL and 15/20/30 mg/dL if glucose value ≤100 mg/dL. Matched paired glucose values ( = 612) from 36 patients (median age 14 years, 58.3% non-Hispanic White, 47.2% male) and 42 inpatient encounters were included in this subanalysis of DKA admissions. The MARDs for DKA and non-DKA admissions ( = 503) were 11.8% and 11.7%, with 97.6% and 98.6% of pairs falling within A and B zones of the Clarke Error Grid, respectively. Severe DKA admissions (pH <7.15 and/or bicarbonate <5 mmol/L) had a MARD of 8.9% compared to 14.3% for nonsevere DKA admissions. The MARD during administration of IV insulin ( = 266) was 13.4%. CGM accuracy is similar between DKA and non-DKA admissions and is maintained in severe DKA and during IV insulin administration, suggesting potential usability in pediatric hospitalizations. Further study on the feasibility of implementation of CGM in the hospital is needed.

Citing Articles

Expanding the horizon of continuous glucose monitoring into the future of pediatric medicine.

Morales-Dopico L, MacLeish S Pediatr Res. 2024; 96(6):1464-1474.

PMID: 39306610 PMC: 11624137. DOI: 10.1038/s41390-024-03573-x.

References
1.
Ekhlaspour L, Mondesir D, Lautsch N, Balliro C, Hillard M, Magyar K . Comparative Accuracy of 17 Point-of-Care Glucose Meters. J Diabetes Sci Technol. 2016; 11(3):558-566. PMC: 5505415. DOI: 10.1177/1932296816672237. View

2.
Cobry E, Pyle L, Waterman L, Forlenza G, Towers L, Karami A . Accuracy of a Continuous Glucose Monitor During Pediatric Type 1 Diabetes Inpatient Admissions. Diabetes Technol Ther. 2024; 26(2):119-124. PMC: 11535460. DOI: 10.1089/dia.2023.0375. View

3.
Rewers A, Dong F, Slover R, Klingensmith G, Rewers M . Incidence of diabetic ketoacidosis at diagnosis of type 1 diabetes in Colorado youth, 1998-2012. JAMA. 2015; 313(15):1570-2. DOI: 10.1001/jama.2015.1414. View

4.
Tzimenatos L, Nigrovic L . Managing Diabetic Ketoacidosis in Children. Ann Emerg Med. 2021; 78(3):340-345. DOI: 10.1016/j.annemergmed.2021.02.028. View

5.
Buschur E, Faulds E, Dungan K . CGM in the Hospital: Is It Ready for Prime Time?. Curr Diab Rep. 2022; 22(9):451-460. PMC: 9261155. DOI: 10.1007/s11892-022-01484-x. View