The Performance and Usability of a Factory-Calibrated Flash Glucose Monitoring System

Overview
Date 2015 Jul 15
PMID 26171659
Citations 301
Authors
Affiliations
Soon will be listed here.
Abstract

Introduction: The purpose of the study was to evaluate the performance and usability of the FreeStyle(®) Libre™ Flash glucose monitoring system (Abbott Diabetes Care, Alameda, CA) for interstitial glucose results compared with capillary blood glucose results.

Materials And Methods: Seventy-two study participants with type 1 or type 2 diabetes were enrolled by four U.S. clinical sites. A sensor was inserted on the back of each upper arm for up to 14 days. Three factory-only calibrated sensor lots were used in the study. Sensor glucose measurements were compared with capillary blood glucose (BG) results (approximately eight per day) obtained using the BG meter built into the reader (BG reference) and with the YSI analyzer (Yellow Springs Instrument, Yellow Springs, OH) reference tests at three clinic visits (32 samples per visit). Sensor readings were masked to the participants.

Results: The accuracy of the results was demonstrated against capillary BG reference values, with 86.7% of sensor results within Consensus Error Grid Zone A. The percentage of readings within Consensus Error Grid Zone A on Days 2, 7, and 14 was 88.4%, 89.2%, and 85.2%, respectively. The overall mean absolute relative difference was 11.4%. The mean lag time between sensor and YSI reference values was 4.5±4.8 min. Sensor accuracy was not affected by factors such as body mass index, age, type of diabetes, clinical site, insulin administration, or hemoglobin A1c.

Conclusions: Interstitial glucose measurements with the FreeStyle Libre system were found to be accurate compared with capillary BG reference values, with accuracy remaining stable over 14 days of wear and unaffected by patient characteristics.

Citing Articles

Effectiveness of a low-glycaemic-index formula on post-gastrectomy hypoglycaemia in patients with gastric cancer: randomized crossover study.

Kubota T, Ohashi T, Nishibeppu K, Takabatake K, Inoue H, Nakabayashi Y BJS Open. 2025; 9(2).

PMID: 40067681 PMC: 11895506. DOI: 10.1093/bjsopen/zraf001.


International Society for Pediatric and Adolescent Diabetes Clinical Practice Consensus Guidelines 2024 Diabetes Technologies: Glucose Monitoring.

Tauschman M, Cardona-Hernandez R, DeSalvo D, Hood K, Laptev D, Lindholm Olinder A Horm Res Paediatr. 2025; 97(6):615-635.

PMID: 39884260 PMC: 11854985. DOI: 10.1159/000543156.


Accuracy of Freestyle Libre continuous glucose monitoring system in critically ill patients after cardiac surgery.

Gu J, Zhao Z, Li H, Li B, Chen S, Cao Y J Cardiothorac Surg. 2025; 20(1):75.

PMID: 39833920 PMC: 11749162. DOI: 10.1186/s13019-024-03229-1.


Accuracy and feasibility of continuous glucose monitoring system in pancreatectomy patients.

Choi Y, Jeon S, Yu S, Jo H, Kim D, Yu Y Langenbecks Arch Surg. 2025; 410(1):32.

PMID: 39792174 DOI: 10.1007/s00423-024-03601-8.


Effects of Synbiotic Supplementation on Bone and Metabolic Health in Caucasian Postmenopausal Women: Rationale and Design of the OsteoPreP Trial.

Turbic A, Vandenput L, Gandham A, Lorentzon M Nutrients. 2024; 16(23).

PMID: 39683612 PMC: 11644401. DOI: 10.3390/nu16234219.


References
1.
Matuleviciene V, Joseph J, Andelin M, Hirsch I, Attvall S, Pivodic A . A clinical trial of the accuracy and treatment experience of the Dexcom G4 sensor (Dexcom G4 system) and Enlite sensor (guardian REAL-time system) tested simultaneously in ambulatory patients with type 1 diabetes. Diabetes Technol Ther. 2014; 16(11):759-67. PMC: 4201312. DOI: 10.1089/dia.2014.0238. View

2.
Thomas L, Kane M, Bakst G, Busch R, Hamilton R, Abelseth J . A glucose meter accuracy and precision comparison: the FreeStyle Flash Versus the Accu-Chek Advantage, Accu-Chek Compact Plus, Ascensia Contour, and the BD Logic. Diabetes Technol Ther. 2008; 10(2):102-10. DOI: 10.1089/dia.2007.0244. View

3.
Kuwa K, Nakayama T, Hoshino T, Tominaga M . Relationships of glucose concentrations in capillary whole blood, venous whole blood and venous plasma. Clin Chim Acta. 2001; 307(1-2):187-92. DOI: 10.1016/s0009-8981(01)00426-0. View

4.
Pfutzner A, Mitri M, Musholt P, Sachsenheimer D, Borchert M, Yap A . Clinical assessment of the accuracy of blood glucose measurement devices. Curr Med Res Opin. 2012; 28(4):525-31. DOI: 10.1185/03007995.2012.673479. View

5.
Cengiz E, Tamborlane W . A tale of two compartments: interstitial versus blood glucose monitoring. Diabetes Technol Ther. 2009; 11 Suppl 1:S11-6. PMC: 2903977. DOI: 10.1089/dia.2009.0002. View