» Articles » PMID: 32211468

Prosthetic Femoral Head Erosion Through an Acetabular Component Treated with Revision and Implant Preservation

Overview
Publisher Elsevier
Date 2020 Mar 27
PMID 32211468
Citations 2
Authors
Affiliations
Soon will be listed here.
Abstract

We present a case report of the rare complication of a femoral head prosthesis eroding through the acetabular liner and shell resulting in a clinical presentation of pseudo-dislocation. The patient presented with a 1-month history of progressive anterior and peritrochanteric hip pain without antecedent trauma. Radiographs demonstrated presumed hip dislocation with superior-posterior superimposition of the femoral head over the acetabular component. The patient underwent revision total hip arthroplasty with intraoperative evidence of extensive metallosis, osteolysis, and femoral head erosion through the acetabular polyethylene liner, acetabular shell, and implantation into the ilium. The femoral stem and acetabular shell were well-fixed, allowing for filling of the defect with bone void filler and cementation of a new polyethylene liner into the acetabular shell.

Citing Articles

Anterior Subluxation of a Metal-on-Metal Total Hip Arthroplasty Resulting in Erosion and Metal Debris.

Park C, Verhey J, Austin R, Howgate D, Bathini A, Lyons M Case Rep Orthop. 2025; 2025:3718607.

PMID: 40017540 PMC: 11867716. DOI: 10.1155/cro/3718607.


The Mechanism of Metallosis After Total Hip Arthroplasty.

Ude C, Esdaille C, Ogueri K, Ho-Man K, Laurencin S, Nair L Regen Eng Transl Med. 2022; 7(3):247-261.

PMID: 35530571 PMC: 9075182. DOI: 10.1007/s40883-021-00222-1.


Do Not Postpone Revision of Worn Conventional Liners in Ceramic-on-Polyethylene Total Hip Arthroplasty: A New Dramatic Failure.

Gehrke T, Citak M, Abdelaziz H Arthroplast Today. 2021; 10:108-113.

PMID: 34345640 PMC: 8319508. DOI: 10.1016/j.artd.2021.06.011.

References
1.
Knox D, Hamilton S, Wardlaw D, Gibson P . Early catastrophic acetabular failure in Furlong total hip replacements. J Orthop Traumatol. 2009; 10(1):39-42. PMC: 2657347. DOI: 10.1007/s10195-008-0045-z. View

2.
Kawamura H, Mishima H, Sugaya H, Nishino T, Shimizu Y, Miyakawa S . The 21- to 27-year results of the Harris-Galante cementless total hip arthroplasty. J Orthop Sci. 2016; 21(3):342-7. DOI: 10.1016/j.jos.2016.02.004. View

3.
Sathappan S, Wee J, Ginat D, Meere P . Massive wear and metallosis of an Acetabular Cup System presenting as pseudodislocation. Orthopedics. 2009; 32(6):449. DOI: 10.3928/01477447-20090511-23. View

4.
Valle A, Ruzo P, Li S, Pellicci P, Sculco T, Salvati E . Dislodgment of polyethylene liners in first and second-generation Harris-Galante acetabular components. A report of eighteen cases. J Bone Joint Surg Am. 2001; 83(4):553-9. DOI: 10.2106/00004623-200104000-00010. View

5.
Young G, Abdel M, Amendola R, Goetz D, Lewallen D, Callaghan J . Cementing Constrained Liners Into Secure Cementless Shells: A Minimum 15-Year Follow-Up Study. J Arthroplasty. 2017; 32(11):3480-3483. PMC: 6322832. DOI: 10.1016/j.arth.2017.07.014. View