» Articles » PMID: 18514888

Catastrophic Failure of Ceramic-polyethylene Bearing Total Hip Arthroplasty

Overview
Journal J Arthroplasty
Specialty Orthopedics
Date 2008 Jun 3
PMID 18514888
Citations 11
Authors
Affiliations
Soon will be listed here.
Abstract

Complications of ceramic-polyethylene bearing total hip arthroplasty (THA) include osteolysis, loosening, dislocation, and component failure. Catastrophic acetabular component failure involves severe damage to both the polyethylene liner and metal shell. This case study presents the first reported complete wear-through of the acetabular portion of a ceramic-polyethylene arthroplasty presenting as a dislocation and a review of the literature. In this study, a patient's alumina ceramic femoral head penetrated the polyethylene liner and titanium shell and presented as a dislocated THA. The contributing factors for this catastrophic failure include young patient age, high activity level, thin polyethylene liner, backside wear, component positioning, polyethylene sterilization with gamma irradiation in air, and lack of appropriate follow-up. Revision THA was performed without complications.

Citing Articles

Eccentric Wear of Polyethylene Liner in a Bipolar Hemiarthroplasty Leading to a Revision Total Hip Arthroplasty.

Archik S, Angachekar D, Kataria J, Nakade S J Orthop Case Rep. 2024; 14(8):71-75.

PMID: 39157471 PMC: 11327670. DOI: 10.13107/jocr.2024.v14.i08.4650.


[Particle disease and its effects on periarticular tissue].

Donner S, Perka C, Krenn V, Ort M Orthopadie (Heidelb). 2023; 52(3):196-205.

PMID: 36867226 DOI: 10.1007/s00132-023-04348-8.


Do Not Postpone Revision of Worn Conventional Liners in Ceramic-on-Polyethylene Total Hip Arthroplasty: A New Dramatic Failure.

Gehrke T, Citak M, Abdelaziz H Arthroplast Today. 2021; 10:108-113.

PMID: 34345640 PMC: 8319508. DOI: 10.1016/j.artd.2021.06.011.


Catastrophic failure and metallosis of the acetabular component in total hip arthroplasty.

Ozan F, Kahraman M, Baktir A, Gencer K J Orthop Surg Res. 2021; 16(1):349.

PMID: 34051798 PMC: 8164279. DOI: 10.1186/s13018-021-02492-5.


Catastrophic wear in uncemented acetabular cups after Total Hip Arthroplasty. A case series.

Jasqui Remba S, Esquivel Gomez R, Gutierrez Ambriz L, Portman Santos D, Baley Amiga I Int J Surg Case Rep. 2021; 81:105846.

PMID: 33887869 PMC: 8049991. DOI: 10.1016/j.ijscr.2021.105846.