» Articles » PMID: 30083199

Cyclic Fatigue Resistance of Heat-Treated Nickel-Titanium Instruments

Overview
Journal Iran Endod J
Date 2018 Aug 8
PMID 30083199
Citations 16
Authors
Affiliations
Soon will be listed here.
Abstract

Introduction: This study compared the cyclic fatigue resistance (CFR) of new instruments manufactured by heat-treated nickel-titanium wire.

Methods And Materials: Ninety-six new instruments from HyFlex CM (HF), Edge File (EF), Pro Design S (PDS/L) and Mtwo (MT) (20/0.06 and 25/0.06) (=12) systems were evaluated. A stainless steel device was used and time and number of cycles to fracture (NCF) were observed. Fragments were measured and fracture surface was evaluated using scanning electron microscope (SEM). ANOVA and Tukey's tests were applied with the level of significance set at 0.05.

Results: PDS 20/0.06 and PDL 25/0.06 instruments presented the highest CFR. MT 20/0.06 and MT 25/0.06 showed the lowest CFR (<0.05). The length of fragments was similar for 25/0.06 instruments and HF presented the highest one for 20/0.06 instruments. SEM analyses showed morphology suggestive of ductile fracture.

Conclusion: Heat treatment increased resistance to cyclic fatigue differently for each type of instrument. PDS 20/0.06 and PDL 25/0.06 present higher cyclic fatigue resistance.

Citing Articles

Efficacy of Controlled Memory and Shape Memory Nickel Titanium Instruments in Removing Filling Material from Severely Curved Root Canals: An Study.

Zevallos-Quiroz C, Perez I, Garcia-Rupaya C Iran Endod J. 2023; 14(2):115-121.

PMID: 36855448 PMC: 9968390. DOI: 10.22037/iej.v14i2.23644.


Effect of hydrochloric acid, sodium hypochlorite, and autoclave sterilization on the force characteristics of orthodontic NiTi-closed coils.

Noorollahian S, Khaleghi A Dent Res J (Isfahan). 2023; 20:8.

PMID: 36820142 PMC: 9937939.


Anatomic Two-dimensional and Three-dimensional Models for Cyclic Fatigue Testing of Endodontic Instruments.

Piasecki L, Makowka S, Gambarini G Iran Endod J. 2023; 15(2):100-105.

PMID: 36704437 PMC: 9709840. DOI: 10.22037/iej.v15i2.27342.


Comparison of Mechanical and Structural Properties of Nickel-titanium Alloy with Titanium-molybdenum Alloy and Titanium-niobium Alloy as Potential Metals for Endodontic Files.

Nino-Barrera J, Aldana-Ojeda L, Gamboa-Martinez L, Acosta-Humanez M, Silva-Castellanos C, Cortes-Rodriguez C Iran Endod J. 2023; 16(1):49-55.

PMID: 36704415 PMC: 9709876. DOI: 10.22037/iej.v16i1.24048.


Influence of Rotation Speed of Mtwo Files on Root Canal Instrumentation Time with Different Canal Curvatures.

Carlesi T, Plotino G, Colangeli M, Pappalardo A, Malagnino V Iran Endod J. 2023; 16(1):38-42.

PMID: 36704413 PMC: 9709877. DOI: 10.22037/iej.v16i1.24773.


References
1.
Bulem U, Kececi A, Guldas H . Experimental evaluation of cyclic fatigue resistance of four different nickel-titanium instruments after immersion in sodium hypochlorite and/or sterilization. J Appl Oral Sci. 2014; 21(6):505-10. PMC: 3891273. DOI: 10.1590/1679-775720130083. View

2.
Braga L, Faria Silva A, Buono V, de Azevedo Bahia M . Impact of heat treatments on the fatigue resistance of different rotary nickel-titanium instruments. J Endod. 2014; 40(9):1494-7. DOI: 10.1016/j.joen.2014.03.007. View

3.
Versiani M, Carvalho K, Mazzi-Chaves J, Sousa-Neto M . Micro-computed Tomographic Evaluation of the Shaping Ability of XP-endo Shaper, iRaCe, and EdgeFile Systems in Long Oval-shaped Canals. J Endod. 2017; 44(3):489-495. DOI: 10.1016/j.joen.2017.09.008. View

4.
Gambarini G, Grande N, Plotino G, Somma F, Garala M, De Luca M . Fatigue resistance of engine-driven rotary nickel-titanium instruments produced by new manufacturing methods. J Endod. 2008; 34(8):1003-5. DOI: 10.1016/j.joen.2008.05.007. View

5.
Plotino G, Testarelli L, Al-Sudani D, Pongione G, Grande N, Gambarini G . Fatigue resistance of rotary instruments manufactured using different nickel-titanium alloys: a comparative study. Odontology. 2012; 102(1):31-5. DOI: 10.1007/s10266-012-0088-8. View