» Articles » PMID: 22414840

An in Vitro Cyclic Fatigue Analysis of Different Endodontic Nickel-titanium Rotary Instruments

Overview
Journal J Endod
Publisher Elsevier
Specialty Dentistry
Date 2012 Mar 15
PMID 22414840
Citations 12
Authors
Affiliations
Soon will be listed here.
Abstract

Introduction: The study compared Twisted File (TF) with 3 traditionally manufactured systems to determine whether changes in the manufacturing process improved the cyclic fatigue resistance.

Methods: Four rotary file systems, (1) ProFile (PF), (2) Mtwo, (3) K3, and (4) TF, were tested in artificial canals with 45° and 90° angles of curvature. Ten instruments each of the 4 file systems were tested in both angles of curvature (n = 10). All instruments had identical size and taper (.06/0.25 tip diameter).

Results: A statistically significant difference (P < .05) was noted between TF and other nickel-titanium instruments in both 45° and 90° angles of curvature. TF showed the greatest mean number of cycles to failure. There was no statistical difference between PF and K3 (P > .05) in both canal curvatures; however, statistically significant difference (P < .05) was observed between Mtwo and the other 2 traditionally manufactured instruments. Mtwo showed the lowest mean number of cycles to failure.

Conclusions: Under the conditions of this study, size .06/0.25 TF was significantly more resistant to fatigue than the other 3 instrument systems produced with the traditional grinding process.

Citing Articles

Comparison of hand, rotary, reciprocating and self-adjusting files in cleaning mesio-buccal canals of maxillary first molars.

Adsare S, Tehreem A, V S, Shetty A, Siraparapu K, Jain L Bioinformation. 2024; 20(7):744-747.

PMID: 39309568 PMC: 11414327. DOI: 10.6026/973206300200744.


The effect of canal curvature on cyclic fatigue resistance of rotary instruments using different irrigation materials ( study).

Alsunboli M, Ihsan S, Sabah D F1000Res. 2024; 12:449.

PMID: 38721249 PMC: 11078266. DOI: 10.12688/f1000research.130249.2.


Upper First and Second Molar Pulp Chamber Endodontic Anatomy Evaluation According to a Recent Classification: A Cone Beam Computed Tomography Study.

Reda R, Di Nardo D, Zanza A, Bellanova V, Abbagnale R, Pagnoni F J Imaging. 2024; 10(1).

PMID: 38248994 PMC: 10817241. DOI: 10.3390/jimaging10010009.


Phase transformation and mechanical properties of heat-treated nickel-titanium rotary endodontic instruments at room and body temperatures.

Kasuga Y, Kimura S, Maki K, Unno H, Omori S, Hirano K BMC Oral Health. 2023; 23(1):825.

PMID: 37904159 PMC: 10614384. DOI: 10.1186/s12903-023-03550-6.


Effect of autoclave sterilization on the cyclic fatigue resistance of EdgeFile X7, 2Shape, and F-one nickel-titanium endodontic instruments.

Al-Amidi A, Al-Gharrawi H J Conserv Dent. 2023; 26(1):26-30.

PMID: 36908740 PMC: 10003289. DOI: 10.4103/jcd.jcd_611_21.