» Articles » PMID: 26372255

Effect of Autoclave Sterilization on the Cyclic Fatigue Resistance of Thermally Treated Nickel-Titanium Instruments

Overview
Journal Int Endod J
Specialty Dentistry
Date 2015 Sep 16
PMID 26372255
Citations 22
Authors
Affiliations
Soon will be listed here.
Abstract

Aim: To compare the cyclic fatigue resistance of HyFlex CM, Twisted Files (TF), K3XF, Race, and K3, and evaluate the effect of autoclave sterilization on the cyclic fatigue resistance of these instruments both before and after the files were cycled.

Methodology: Five types of NiTi instruments with similar size 30, .06 taper were selected: HyFlex CM, TF, K3XF, Race and K3. Files were tested in a simulated canal with a curvature of 60° and a radius of 3 mm. The number of cycles to failure of each instrument was determined to evaluate cyclic fatigue resistance. Each type of instruments was randomly divided into four experimental groups: group 1 (n = 20), unsterilized instruments; group 2 (n = 20), pre-sterilized instruments subjected to 10 cycles of autoclave sterilization; group 3 (n = 20), instruments tested were sterilized at 25%, 50% and 75% of the mean cycles to failure as determined in group 1, and then cycled to failure; group 4 (n = 20), instruments cycled in the same manner as group 3 but without sterilization. The fracture surfaces of instruments were examined by scanning electron microscopy (SEM).

Results: HyFlex CM, TF and K3XF had significantly higher cyclic fatigue resistance than Race and K3 in the unsterilized group 1 (P < 0.05). Autoclave sterilization significantly increased the MCF of HyFlex CM and K3XF (P < 0.05) both before and after the files were cycled. SEM examination revealed a typical pattern of cyclic fatigue fracture in all instruments.

Conclusions: HyFlex CM, TF and K3XF instruments composed of new thermal-treated alloy were more resistant to fatigue failure than Race and K3. Autoclaving extended the cyclic fatigue life of HyFlex CM and K3XF.

Citing Articles

An comparative evaluation of cyclic fatigue resistance of two rotary and two reciprocating file systems.

Srikumar G, Gadbail V, Alexander A, Nishad G, Rahane S, Beautlin J J Conserv Dent Endod. 2024; 27(7):774-779.

PMID: 39262602 PMC: 11385906. DOI: 10.4103/JCDE.JCDE_2_24.


Effect of simulated clinical use and sterilization on the cyclic fatigue resistance of nickel titanium files.

Alajemi M, AbuMostafa A PeerJ. 2024; 12:e17418.

PMID: 38799069 PMC: 11128215. DOI: 10.7717/peerj.17418.


Cutting efficiency and corrosion resistance of heat-treated endodontic files after various disinfection protocols.

Ulusoy S, Basturk F, Turkaydin D, Garip Berker Y, Gunday M, Durmazpinar P Odontology. 2024; 112(3):847-854.

PMID: 38381265 DOI: 10.1007/s10266-023-00896-9.


Effect from Autoclave Sterilization and Usage on the Fracture Resistance of Heat-Treated Nickel-Titanium Rotary Files.

Abed R, Al Raeesi D, Alshehhi A, Alkhatib Z, Khamis A, Jamal M Materials (Basel). 2023; 16(6).

PMID: 36984141 PMC: 10056662. DOI: 10.3390/ma16062261.


Effects of Autoclave Sterilization on Cyclic Fatigue Resistance in 5 Types of Rotary Endodontic Instruments: An In Vitro Study.

Stosic N, Popovic J, Apostolovic M, Mitic A, Nikolic M, Barac R Med Sci Monit. 2023; 29:e939694.

PMID: 36967567 PMC: 10064810. DOI: 10.12659/MSM.939694.