Effect of Differences in Pixel Size on Image Characteristics of Digital Intraoral Radiographic Systems: a Physical and Visual Evaluation

Overview
Date 2020 Apr 18
PMID 32302229
Citations 3
Authors
Affiliations
Soon will be listed here.
Abstract

Objectives: To quantify and validate the effect of pixel size on a digital intraoral radiographic system according to International Electrotechnical Commission standards through physical and visual evaluations.

Methods: The digital intraoral radiographic system used was the photostimulable phosphor imaging plate and scanner system. The system had three image capture modes: high-speed (HS), high-resolution (HR), and super high-resolution (SHR) with different pixels. The physical characteristics of the system were evaluated using presampled modulation transfer function (MTF) and the normalized noise power spectrum (NNPS). An aluminum (Al) step phantom with different depths of holes was used to acquire images under various exposure conditions. The average number of perceptible holes from all steps was plotted against each exposure dose. The results were compared to analyze the effects of pixel size on image quality of intraoral radiographs.

Results: The MTF was slightly higher with SHR than with HR and HS. The NNPS with SHR showed about a 40% decrease in magnitude compared to HS. The total number of perceptible holes in the Al step phantom was higher with SHR than with HS and HR in all exposure conditions.

Conclusions: The MTF and NNPS obtained with different pixel size could be quantified by physical evaluation, and the differences were visually validated with Al step phantom. The SHR mode has the potential to decrease the radiation dose without compromising the image quality.

Citing Articles

How Do Different Image Modules Impact the Accuracy of Working Length Measurements in Digital Periapical Radiography? An In Vitro Study.

Abat V, Kaptan R Diagnostics (Basel). 2025; 15(3).

PMID: 39941235 PMC: 11817009. DOI: 10.3390/diagnostics15030305.


Evaluation of resolution characteristics of digital intraoral radiographic images using a task transfer function.

Kuramoto T, Takarabe S, Tsuru H, Shibayama Y, Kato T, Yoshiura K Oral Radiol. 2022; 38(4):638-644.

PMID: 35767125 DOI: 10.1007/s11282-022-00633-y.


Effects of differences in pixel size on image characteristics of digital intraoral radiographic systems: a physical and visual evaluation.

Watanabe H, Kurabayashi T Dentomaxillofac Radiol. 2021; 51(3):20210093.

PMID: 34914543 PMC: 8925879. DOI: 10.1259/dmfr.20210093.


Effect of beam quality and readout direction in the edge profile on the modulation transfer function of photostimulable phosphor systems via the edge method.

Takarabe S, Kuramoto T, Shibayama Y, Tsuru H, Tatsumi M, Kato T J Med Imaging (Bellingham). 2021; 8(4):043501.

PMID: 34307736 PMC: 8291077. DOI: 10.1117/1.JMI.8.4.043501.

References
1.
Tanaka N, Morishita J, Tsuda N, Ohki M . Effect of readout direction in the edge profile on the modulation transfer function of computed radiographic systems by use of the edge method. Radiol Phys Technol. 2013; 6(2):474-9. DOI: 10.1007/s12194-013-0222-5. View

2.
Brennan J . An introduction to digital radiography in dentistry. J Orthod. 2002; 29(1):66-9. DOI: 10.1093/ortho/29.1.66. View

3.
Doi K, Strubler K, Rossmann K . Truncation errors in calculating the MTF of radiographic screen-film systems from the line spread function. Phys Med Biol. 1972; 17(2):241-50. DOI: 10.1088/0031-9155/17/2/010. View

4.
Vandre R, Pajak J, Abdel-Nabi H, Farman T, Farman A . Comparison of observer performance in determining the position of endodontic files with physical measures in the evaluation of dental X-ray imaging systems. Dentomaxillofac Radiol. 2000; 29(4):216-22. DOI: 10.1038/sj/dmfr/4600533. View

5.
Ideguchi T . [The evaluation of physical imaging properties for digital radiography - the resolution property(the slit method)]. Nihon Hoshasen Gijutsu Gakkai Zasshi. 2009; 65(8):1109-17. DOI: 10.6009/jjrt.65.1109. View