Electrocardiogram-gated Coronary CT Angiography Dose Estimates Using ImPACT

Overview
Date 2016 Jul 26
PMID 27455500
Citations 1
Authors
Affiliations
Soon will be listed here.
Abstract

The primary study objective was to assess radiation doses using a modified form of the Imaging Performance Assessment of Computed Tomography (CT) scanner (ImPACT) patient dosimetry for cardiac applications on an Aquilion ONE ViSION Edition scanner, including the Ca score, target computed tomography angiography (CTA), prospective CTA, continuous CTA/cardiac function analysis (CFA), and CTA/CFA modulation. Accordingly, we clarified the CT dose index (CTDI) to determine the relationship between heart rate (HR) and X-ray exposure. As a secondary objective, we compared radiation doses using modified ImPACT, a whole-body dosimetry phantom study, and the k-factor method to verify the validity of the dose results obtained with modified ImPACT. The effective dose determined for the reference person (4.66 mSv at 60 beats per minute (bpm) and 33.43 mSv at 90bpm) were approximately 10% less than those determined for the phantom study (5.28 mSv and 36.68 mSv). The effective doses according to the k-factor (0.014 mSv•mGy-1•cm-1; 2.57 mSv and 17.10 mSv) were significantly lower than those obtained with the other two methods. In the present study, we have shown that ImPACT, when modified for cardiac applications, can assess both absorbed and effective doses. The results of our dose comparison indicate that modified ImPACT dose assessment is a promising and practical method for evaluating coronary CTA.

Citing Articles

Evaluation of radiation dose for inferior vena cava filter placement during pregnancy: A comparison of dosimetry and dose calculation software.

Matsunaga Y, Haba T, Kobayashi M, Suzuki S, Asada Y, Chida K J Appl Clin Med Phys. 2022; 24(2):e13884.

PMID: 36546565 PMC: 9924124. DOI: 10.1002/acm2.13884.

References
1.
Brownell G, ELLETT W, Reddy A . Absorbed fractions for photon dosimetry. J Nucl Med. 1968; :Suppl :29-39. View

2.
. The 2007 Recommendations of the International Commission on Radiological Protection. ICRP publication 103. Ann ICRP. 2007; 37(2-4):1-332. DOI: 10.1016/j.icrp.2007.10.003. View

3.
Halliburton S, Abbara S, Chen M, Gentry R, Mahesh M, Raff G . SCCT guidelines on radiation dose and dose-optimization strategies in cardiovascular CT. J Cardiovasc Comput Tomogr. 2011; 5(4):198-224. PMC: 3391026. DOI: 10.1016/j.jcct.2011.06.001. View

4.
Kobayashi M, Asada Y, Matsubara K, Matsunaga Y, Kawaguchi A, Katada K . Evaluation of organ doses and effective dose according to the ICRP Publication 110 reference male/female phantom and the modified ImPACT CT patient dosimetry. J Appl Clin Med Phys. 2014; 15(5):4823. PMC: 5711082. DOI: 10.1120/jacmp.v15i5.4823. View

5.
Abbara S, Arbab-Zadeh A, Callister T, Desai M, Mamuya W, Thomson L . SCCT guidelines for performance of coronary computed tomographic angiography: a report of the Society of Cardiovascular Computed Tomography Guidelines Committee. J Cardiovasc Comput Tomogr. 2009; 3(3):190-204. DOI: 10.1016/j.jcct.2009.03.004. View