» Articles » PMID: 27370946

Flow Diverter Effect of LVIS Stent on Cerebral Aneurysm Hemodynamics: a Comparison with Enterprise Stents and the Pipeline Device

Overview
Journal J Transl Med
Publisher Biomed Central
Date 2016 Jul 3
PMID 27370946
Citations 93
Authors
Affiliations
Soon will be listed here.
Abstract

Background: The aim of this study was to quantify the effect of the new Low-profile Visualized Intraluminal Support (LVIS®D) device and the difference of fluid diverting effect compared with the Pipeline device and the Enterprise stent using computational fluid dynamics (CFD).

Methods: In this research, we simulated three aneurysms constructed from 3D digital subtraction angiography (DSA). The Enterprise, LVIS and the Pipeline device were virtually conformed to fit into the vessel lumen and placed across the aneurysm orifice. Computational fluid dynamics analysis was performed to compare the hemodynamic differences such as WSS, Velocity and Pressure among these stents.

Results: Control referred to the unstented model, the percentage of hemodynamic changes were all compared to Control. A single LVIS stent caused more wall shear stress reduction than double Enterprise stents (39.96 vs. 30.51 %) and velocity (23.13 vs. 18.64 %). Significant reduction in wall shear stress (63.88 %) and velocity (46.05 %) was observed in the double-LVIS stents. A single Pipeline showed less reduction in WSS (51.08 %) and velocity (37.87 %) compared with double-LVIS stent. The double-Pipeline stents resulted in the most reduction in WSS (72.37 %) and velocity (54.26 %). Moreover, the pressure increased with minuscule extent after stenting, compared with the unstented model.

Conclusions: This is the first study analyzing flow modifications associated with LVIS stents. We found that the LVIS stent has certain hemodynamic effects on cerebral aneurysms: a single LVIS stent caused more flow reductions than the double-Enterprise stent but less than a Pipeline device. Nevertheless, the double-LVIS stent resulted in a better flow diverting effect than a Pipeline device.

Citing Articles

Analysis of the efficacy and factors affecting the prognosis of intracranial dissecting aneurysm treated with multi-stent overlap technique.

Li X, Hu H, Wang L, Yuan G, Fu C, Ma J BMC Neurol. 2024; 24(1):495.

PMID: 39736529 PMC: 11687087. DOI: 10.1186/s12883-024-03991-8.


Y Stent-Assisted Coiling Technique for Bifurcation Aneurysms Using Double Neuroform® Stent: a Large Restrospective Series.

Caamano I, Remollo S, Terceno M, Blanco A, Bashir S, Castano C Clin Neuroradiol. 2024; 34(4):919-928.

PMID: 39023542 DOI: 10.1007/s00062-024-01437-9.


Appraisal of the Flow Diversion Effect Provided by Braided Intracranial Stents.

Cay F, Arat A J Clin Med. 2024; 13(12).

PMID: 38929937 PMC: 11204822. DOI: 10.3390/jcm13123409.


Endovascular Treatment of Intracranial Aneurysm: The Importance of the Rheological Model in Blood Flow Simulations.

Boniforti M, Vittucci G, Magini R Bioengineering (Basel). 2024; 11(6).

PMID: 38927758 PMC: 11200932. DOI: 10.3390/bioengineering11060522.


A clinical comparative study of dual LVIS devices and single flow diversion stenting for the treatment of unruptured V3-V4 vertebral artery dissection.

Long S, Shi S, Zhang Z, Tian Q, Wei Z, Ma J Interv Neuroradiol. 2024; :15910199241254864.

PMID: 38751187 PMC: 11569745. DOI: 10.1177/15910199241254864.


References
1.
Geyik S, Yavuz K, Yurttutan N, Saatci I, Cekirge H . Stent-assisted coiling in endovascular treatment of 500 consecutive cerebral aneurysms with long-term follow-up. AJNR Am J Neuroradiol. 2013; 34(11):2157-62. PMC: 7964833. DOI: 10.3174/ajnr.A3574. View

2.
Benitez R, Silva M, Klem J, Veznedaroglu E, Rosenwasser R . Endovascular occlusion of wide-necked aneurysms with a new intracranial microstent (Neuroform) and detachable coils. Neurosurgery. 2004; 54(6):1359-67. DOI: 10.1227/01.neu.0000124484.87635.cd. View

3.
Wakhloo A, Tio F, Lieber B, Schellhammer F, Graf M, Hopkins L . Self-expanding nitinol stents in canine vertebral arteries: hemodynamics and tissue response. AJNR Am J Neuroradiol. 1995; 16(5):1043-51. PMC: 8337802. View

4.
Cebral J, Castro M, Putman C, Alperin N . Flow-area relationship in internal carotid and vertebral arteries. Physiol Meas. 2008; 29(5):585-94. PMC: 2692290. DOI: 10.1088/0967-3334/29/5/005. View

5.
Cho Y, Sohn C, Kang H, Kim J, Cho W, Hwang G . Coil embolization of intracranial saccular aneurysms using the Low-profile Visualized Intraluminal Support (LVIS™) device. Neuroradiology. 2014; 56(7):543-51. DOI: 10.1007/s00234-014-1363-x. View