» Articles » PMID: 26585260

Endovascular Treatment of Intracranial Aneurysms with the WEB Device: A Systematic Review of Clinical Outcomes

Overview
Specialty Neurology
Date 2015 Nov 21
PMID 26585260
Citations 29
Authors
Affiliations
Soon will be listed here.
Abstract

Background And Purpose: Despite the improvement in technology, endovascular treatment of bifurcation intracranial wide-neck aneurysms remains challenging, mainly due to the difficulty of maintaining coils within the aneurysm sac without compromising the patency of bifurcation arteries. The Woven EndoBridge (WEB) device is a recent intrasaccular braided device specifically dedicated to treating such aneurysms with a wide neck by disrupting the flow in the aneurysmal neck and promoting progressive aneurysmal thrombosis.

Materials And Methods: Using several health data bases, we conducted a systematic review of all published studies of WEB endovascular treatment in intracranial aneurysms from 2010 onward to evaluate its efficacy and safety profile.

Results: The literature search identified 6 relevant studies (7 articles) including wide-neck bifurcation aneurysms in ≥80% of cases. Clinical data supporting the efficacy and safety of the WEB are limited to noncomparative cohort studies with large heterogeneity from a methodologic standpoint. The WEB deployment was feasible with a success rate of 93%-100%. Permanent morbidity (mRS of >1 at last follow-up) and mortality were measured at 2.2%-6.7% and 0%-17%, respectively. The adequate occlusion rate (total occlusion or neck remnant) varied between 65% and 85.4% at midterm follow-up (range, 3.3-27.4 months).

Conclusions: Endovascular treatment of bifurcation wide-neck aneurysms with the WEB device is feasible and allows an acceptably adequate aneurysm occlusion rate; however, the rate of neck remnants is not negligible. The WEB device needs further clinical and anatomic evaluation with long-term prospective studies, especially of the risk of WEB compression. Prospective controlled studies should be encouraged.

Citing Articles

Intrasaccular Treatment of Intracranial Aneurysms: A Comprehensive Review.

Shao M, White T, Bassett J, Dowlati E, Mehta S, Werner C J Clin Med. 2024; 13(20).

PMID: 39458111 PMC: 11508718. DOI: 10.3390/jcm13206162.


Comments on "treatment of large intracranial aneurysms using woven endobridge".

Fatima Y, Qasim M, Magsj I Neurosurg Rev. 2024; 47(1):474.

PMID: 39181995 DOI: 10.1007/s10143-024-02730-4.


Dual Layer vs Single Layer Woven EndoBridge Device in the Treatment of Intracranial Aneurysms: A Propensity Score-Matched Analysis.

Dmytriw A, Salim H, Musmar B, Aslan A, Cancelliere N, McLellan R Neurosurg Rev. 2024; 47(1):116.

PMID: 38483647 DOI: 10.1007/s10143-024-02341-z.


[Brain aneurysms treatment with pCONus2 device. Initial experience at IMSS].

Santos-Franco J, Cruz-Arguelles C, Abrego-Salinas A, Casas-Martinez M, DeFont-Reaulx Rojas E Rev Med Inst Mex Seguro Soc. 2023; 61(3):265-273.

PMID: 37216405 PMC: 10437233.


Quantitative angiographic haemodynamic evaluation of bypasses for complex aneurysms: a preliminary study.

Lu J, Xue C, Hu X, Zhao Y, Zhang D, Chen X Stroke Vasc Neurol. 2021; 7(1):54-61.

PMID: 34642254 PMC: 8899645. DOI: 10.1136/svn-2021-000858.


References
1.
Gory B, Aguilar-Perez M, Pomero E, Turjman F, Weber W, Fischer S . pCONus Device for the Endovascular Treatment of Wide-Neck Middle Cerebral Artery Aneurysms. AJNR Am J Neuroradiol. 2015; 36(9):1735-40. PMC: 7968775. DOI: 10.3174/ajnr.A4392. View

2.
Liebig T, Kabbasch C, Strasilla C, Berlis A, Weber W, Pierot L . Intrasaccular Flow Disruption in Acutely Ruptured Aneurysms: A Multicenter Retrospective Review of the Use of the WEB. AJNR Am J Neuroradiol. 2015; 36(9):1721-7. PMC: 7968782. DOI: 10.3174/ajnr.A4347. View

3.
Moher D, Liberati A, Tetzlaff J, Altman D . Preferred reporting items for systematic reviews and meta-analyses: the PRISMA statement. PLoS Med. 2009; 6(7):e1000097. PMC: 2707599. DOI: 10.1371/journal.pmed.1000097. View

4.
Raymond J, Guilbert F, Weill A, Georganos S, Juravsky L, Lambert A . Long-term angiographic recurrences after selective endovascular treatment of aneurysms with detachable coils. Stroke. 2003; 34(6):1398-403. DOI: 10.1161/01.STR.0000073841.88563.E9. View

5.