Correcting for Respiratory Motion in Liver PET/MRI: Preliminary Evaluation of the Utility of Bellows and Navigated Hepatobiliary Phase Imaging

Overview
Journal EJNMMI Phys
Specialty Radiology
Date 2015 Oct 27
PMID 26501822
Citations 8
Authors
Affiliations
Soon will be listed here.
Abstract

Background: The purpose of this study was to evaluate the utility of bellows-based respiratory compensation and navigated hepatobiliary phase imaging to correct for respiratory motion in the setting of dedicated liver PET/MRI.

Methods: Institutional review board approval and informed consent were obtained. Six patients with metastatic neuroendocrine tumor were imaged using Ga-68 DOTA-TOC PET/MRI. Whole body imaging and a dedicated 15-min liver PET acquisition was performed, in addition to navigated and breath-held hepatobiliary phase (HBP) MRI. Liver PET data was reconstructed three ways: the entire data set (liver PET), gated using respiratory bellows (RC-liver PET), and a non-gated data set reconstructed using the same amount of data used in the RC-liver PET (shortened liver PET). Liver lesions were evaluated using SUVmax, SUVpeak, SUVmean, and Volisocontour. Additionally, the displacement of each lesion between the RC-liver PET images and the navigated and breath-held HBP images was calculated.

Results: Respiratory compensation resulted in a 43 % increase in SUVs compared to ungated data (liver vs RC-liver PET SUVmax 26.0 vs 37.3, p < 0.001) and a 25 % increase compared to a non-gated reconstruction using the same amount of data (RC-liver vs shortened liver PET SUVmax 26.0 vs 32.6, p < 0.001). Lesion displacement was minimized using navigated HBP MRI (1.3 ± 1.0 mm) compared to breath-held HBP MRI (23.3 ± 1.0 mm).

Conclusions: Respiratory bellows can provide accurate respiratory compensation when imaging liver lesions using PET/MRI, and results in increased SUVs due to a combination of increased image noise and reduced respiratory blurring. Additionally, navigated HBP MRI accurately aligns with respiratory compensated PET data.

Citing Articles

Potential applications of PET/MRI in non-oncologic conditions within the abdomen and pelvis.

Bartlett D, Takahashi H, Bach C, Lunn B, Thorpe M, Broski S Abdom Radiol (NY). 2023; 48(12):3624-3633.

PMID: 37145312 DOI: 10.1007/s00261-023-03922-0.


Impact of low injected activity on data driven respiratory gating for PET/CT imaging with continuous bed motion.

Meier J, Diab R, Connor T, Mawlawi O J Appl Clin Med Phys. 2022; 23(5):e13619.

PMID: 35481961 PMC: 9121057. DOI: 10.1002/acm2.13619.


Evaluating two respiratory correction methods for abdominal PET/MRI imaging.

Ruan W, Liu F, Sun X, Hu F, Wu T, Zhang Y EJNMMI Phys. 2022; 9(1):5.

PMID: 35099646 PMC: 8804027. DOI: 10.1186/s40658-022-00430-w.


Ultrasound-based sensors for respiratory motion assessment in multimodality PET imaging.

Madore B, Belsley G, Cheng C, Preiswerk F, Kijewski M, Wu P Phys Med Biol. 2021; 67(2).

PMID: 34891142 PMC: 9482332. DOI: 10.1088/1361-6560/ac4213.


Synergistic motion compensation strategies for positron emission tomography when acquired simultaneously with magnetic resonance imaging.

Polycarpou I, Soultanidis G, Tsoumpas C Philos Trans A Math Phys Eng Sci. 2021; 379(2204):20200207.

PMID: 34218675 PMC: 8255946. DOI: 10.1098/rsta.2020.0207.


References
1.
Manber R, Thielemans K, Hutton B, Barnes A, Ourselin S, Arridge S . Practical PET Respiratory Motion Correction in Clinical PET/MR. J Nucl Med. 2015; 56(6):890-6. DOI: 10.2967/jnumed.114.151779. View

2.
Martinez-Moller A, Eiber M, Nekolla S, Souvatzoglou M, Drzezga A, Ziegler S . Workflow and scan protocol considerations for integrated whole-body PET/MRI in oncology. J Nucl Med. 2012; 53(9):1415-26. DOI: 10.2967/jnumed.112.109348. View

3.
Keller S, Holm S, Hansen A, Sattler B, Andersen F, Klausen T . Image artifacts from MR-based attenuation correction in clinical, whole-body PET/MRI. MAGMA. 2012; 26(1):173-81. DOI: 10.1007/s10334-012-0345-4. View

4.
Barnwell J, Raptis C, McConathy J, Laforest R, Siegel B, Woodard P . Beyond whole-body imaging: advanced imaging techniques of PET/MRI. Clin Nucl Med. 2014; 40(2):e88-95. DOI: 10.1097/RLU.0000000000000567. View

5.
Dawood M, Buther F, Jiang X, Schafers K . Respiratory motion correction in 3-D PET data with advanced optical flow algorithms. IEEE Trans Med Imaging. 2008; 27(8):1164-75. DOI: 10.1109/TMI.2008.918321. View