Respiratory-gated Positron Emission Tomography and Breath-hold Computed Tomography Coupling to Reduce the Influence of Respiratory Motion: Methodology and Feasibility

Overview
Journal Acta Radiol
Specialty Radiology
Date 2008 Dec 20
PMID 19096952
Citations 6
Authors
Affiliations
Soon will be listed here.
Abstract

Background: Respiratory motion causes uptake in positron emission tomography (PET) images of chest and abdominal structures to be blurred and reduced in intensity.

Purpose: To compare two respiratory-gated PET binning methods (based on frequency and amplitude analyses of the respiratory signal) and to propose a "BH-based" method based on an additional breath-hold computed tomography (CT) acquisition.

Material And Methods: Respiratory-gated PET consists in list-mode (LM) acquisition with simultaneous respiratory signal recording. A phantom study featured rectilinear movement of a 0.5-ml sphere filled with (18)F-fluorodeoxyglucose ((18)F-FDG) solution, placed in a radioactive background (sphere-to-background contrast 6:1). Two patients were also examined. Three figures of merit were calculated: the target-to-background ratio profile (TBRP) in the axial direction through the uptake (i.e., the sphere or lesion), full-width-at-half-maximum (FWHM) values, and maximized standard uptake values (SUV(max)).

Results: In the phantom study, the peak TBRP was 0.9 for non-gated volume, 1.83 for BH-based volume, and varied between 1.13 and 1.73 for Freq-based volumes and between 1.34 and 1.66 for Amp-based volumes. A reference volume (REF-static) was also acquired for the phantom (in a static, "expiratory" state), with a peak TBRP at 1.88. TBRPs were computed for patient data, with higher peak values for all gated volumes than for non-gated volumes.

Conclusion: Respiratory-gated PET acquisition reduces the blurring effect and increases image contrast. However, Freq-based and Amp-based volumes are still influenced by inappropriate attenuation correction and misregistration of mobile lesions on CT images. The proposed BH-based method both reduces motion artifacts and improves PET-CT registration.

Citing Articles

Is Prone Position [F]FDG PET/CT Useful in Reducing Respiratory Motion Artifacts in Evaluating Hepatic Lesions?.

Lee C, Son H, Woo J, Lee S Diagnostics (Basel). 2023; 13(15).

PMID: 37568906 PMC: 10417611. DOI: 10.3390/diagnostics13152539.


The impact of data-driven respiratory gating in clinical F-18 FDG PET/CT: comparison of free breathing and deep-expiration breath-hold CT protocol.

Kang S, Moon B, Kim H, Yoon H, Kim B Ann Nucl Med. 2021; 35(3):328-337.

PMID: 33449303 DOI: 10.1007/s12149-020-01574-4.


Clinical respiratory motion correction software (reconstruct, register and averaged-RRA), for F-FDG-PET-CT: phantom validation, practical implications and patient evaluation.

Bouyeure-Petit A, Chastan M, Edet-Sanson A, Becker S, Thureau S, Houivet E Br J Radiol. 2016; 90(1070):20160549.

PMID: 27936893 PMC: 5685101. DOI: 10.1259/bjr.20160549.


Correcting for respiratory motion in liver PET/MRI: preliminary evaluation of the utility of bellows and navigated hepatobiliary phase imaging.

Hope T, Verdin E, Bergsland E, Ohliger M, Corvera C, Nakakura E EJNMMI Phys. 2015; 2(1):21.

PMID: 26501822 PMC: 4573645. DOI: 10.1186/s40658-015-0125-0.


Application of partial volume effect correction and 4D PET in the quantification of FDG avid lung lesions.

Salavati A, Borofsky S, Boon-Keng T, Houshmand S, Khiewvan B, Saboury B Mol Imaging Biol. 2014; 17(1):140-8.

PMID: 25080325 DOI: 10.1007/s11307-014-0776-6.