Motion Compensation for Brain PET Imaging Using Wireless MR Active Markers in Simultaneous PET-MR: Phantom and Non-human Primate Studies

Overview
Journal Neuroimage
Specialty Radiology
Date 2014 Jan 15
PMID 24418501
Citations 18
Authors
Affiliations
Soon will be listed here.
Abstract

Brain PET scanning plays an important role in the diagnosis, prognostication and monitoring of many brain diseases. Motion artifacts from head motion are one of the major hurdles in brain PET. In this work, we propose to use wireless MR active markers to track head motion in real time during a simultaneous PET-MR brain scan and incorporate the motion measured by the markers in the listmode PET reconstruction. Several wireless MR active markers and a dedicated fast MR tracking pulse sequence module were built. Data were acquired on an ACR Flangeless PET phantom with multiple spheres and a non-human primate with and without motion. Motions of the phantom and monkey's head were measured with the wireless markers using a dedicated MR tracking sequence module. The motion PET data were reconstructed using list-mode reconstruction with and without motion correction. Static reference was used as gold standard for quantitative analysis. The motion artifacts, which were prominent on the images without motion correction, were eliminated by the wireless marker based motion correction in both the phantom and monkey experiments. Quantitative analysis was performed on the phantom motion data from 24 independent noise realizations. The reduction of bias of sphere-to-background PET contrast by active marker based motion correction ranges from 26% to 64% and 17% to 25% for hot (i.e., radioactive) and cold (i.e., non-radioactive) spheres, respectively. The motion correction improved the channelized Hotelling observer signal-to-noise ratio of the spheres by 1.2 to 6.9 depending on their locations and sizes. The proposed wireless MR active marker based motion correction technique removes the motion artifacts in the reconstructed PET images and yields accurate quantitative values.

Citing Articles

Effects of List-Mode-Based Intraframe Motion Correction in Dynamic Brain PET Imaging.

Tiss A, Chemli Y, Guehl N, Marin T, Johnson K, El Fakhri G IEEE Trans Radiat Plasma Med Sci. 2024; 8(8):950-958.

PMID: 39507127 PMC: 11540417. DOI: 10.1109/trpms.2024.3432322.


Deep learning applications for quantitative and qualitative PET in PET/MR: technical and clinical unmet needs.

Yang J, Afaq A, Sibley R, McMilan A, Pirasteh A MAGMA. 2024; 37(4):749-763.

PMID: 39167304 DOI: 10.1007/s10334-024-01199-y.


Synergistic motion compensation strategies for positron emission tomography when acquired simultaneously with magnetic resonance imaging.

Polycarpou I, Soultanidis G, Tsoumpas C Philos Trans A Math Phys Eng Sci. 2021; 379(2204):20200207.

PMID: 34218675 PMC: 8255946. DOI: 10.1098/rsta.2020.0207.


Quantitative PET in the 2020s: a roadmap.

Meikle S, Sossi V, Roncali E, Cherry S, Banati R, Mankoff D Phys Med Biol. 2020; 66(6):06RM01.

PMID: 33339012 PMC: 9358699. DOI: 10.1088/1361-6560/abd4f7.


Motion correction for PET data using subspace-based real-time MR imaging in simultaneous PET/MR.

Marin T, Djebra Y, Han P, Chemli Y, Bloch I, El Fakhri G Phys Med Biol. 2020; 65(23):235022.

PMID: 33263317 PMC: 7985095. DOI: 10.1088/1361-6560/abb31d.


References
1.
Tellmann L, Fulton R, Pietrzyk U, Nickel I, Stangier I, Winz O . Concepts of registration and correction of head motion in positron emission tomography. Z Med Phys. 2006; 16(1):67-74. DOI: 10.1078/0939-3889-00293. View

2.
Gifford H, King M, de Vries D, Soares E . Channelized hotelling and human observer correlation for lesion detection in hepatic SPECT imaging. J Nucl Med. 2000; 41(3):514-21. View

3.
Sengupta S, Tadanki S, Gore J, Welch E . Prospective real-time head motion correction using inductively coupled wireless NMR probes. Magn Reson Med. 2013; 72(4):971-85. PMC: 4304771. DOI: 10.1002/mrm.25001. View

4.
Gunn R, Murthy V, Catafau A, Searle G, Bullich S, Slifstein M . Translational characterization of [11C]GSK931145, a PET ligand for the glycine transporter type 1. Synapse. 2011; 65(12):1319-32. DOI: 10.1002/syn.20966. View

5.
Howell L, Murnane K . Nonhuman primate positron emission tomography neuroimaging in drug abuse research. J Pharmacol Exp Ther. 2011; 337(2):324-34. PMC: 3083112. DOI: 10.1124/jpet.108.136689. View