Accelerated Acquisition of Tagged MRI for Cardiac Motion Correction in Simultaneous PET-MR: Phantom and Patient Studies

Overview
Journal Med Phys
Specialty Biophysics
Date 2015 Feb 6
PMID 25652521
Citations 14
Authors
Affiliations
Soon will be listed here.
Abstract

Purpose: Degradation of image quality caused by cardiac and respiratory motions hampers the diagnostic quality of cardiac PET. It has been shown that improved diagnostic accuracy of myocardial defect can be achieved by tagged MR (tMR) based PET motion correction using simultaneous PET-MR. However, one major hurdle for the adoption of tMR-based PET motion correction in the PET-MR routine is the long acquisition time needed for the collection of fully sampled tMR data. In this work, the authors propose an accelerated tMR acquisition strategy using parallel imaging and/or compressed sensing and assess the impact on the tMR-based motion corrected PET using phantom and patient data.

Methods: Fully sampled tMR data were acquired simultaneously with PET list-mode data on two simultaneous PET-MR scanners for a cardiac phantom and a patient. Parallel imaging and compressed sensing were retrospectively performed by GRAPPA and kt-FOCUSS algorithms with various acceleration factors. Motion fields were estimated using nonrigid B-spline image registration from both the accelerated and fully sampled tMR images. The motion fields were incorporated into a motion corrected ordered subset expectation maximization reconstruction algorithm with motion-dependent attenuation correction.

Results: Although tMR acceleration introduced image artifacts into the tMR images for both phantom and patient data, motion corrected PET images yielded similar image quality as those obtained using the fully sampled tMR images for low to moderate acceleration factors (<4). Quantitative analysis of myocardial defect contrast over ten independent noise realizations showed similar results. It was further observed that although the image quality of the motion corrected PET images deteriorates for high acceleration factors, the images were still superior to the images reconstructed without motion correction.

Conclusions: Accelerated tMR images obtained with more than 4 times acceleration can still provide relatively accurate motion fields and yield tMR-based motion corrected PET images with similar image quality as those reconstructed using fully sampled tMR data. The reduction of tMR acquisition time makes it more compatible with routine clinical cardiac PET-MR studies.

Citing Articles

Conditional diffusion-generated super-resolution for myocardial perfusion MRI.

Sun C, Goyal N, Wang Y, Tharp D, Kumar S, Altes T Front Cardiovasc Med. 2025; 12:1499593.

PMID: 39925981 PMC: 11802533. DOI: 10.3389/fcvm.2025.1499593.


Development and first implementation of a novel multi-modality cardiac motion and dosimetry phantom for radiotherapy applications.

Gregg K, Ruff C, Koenig G, Penev K, Shepard A, Kreissler G Med Phys. 2024; 51(10):7479-7491.

PMID: 39042362 PMC: 11798577. DOI: 10.1002/mp.17315.


Highly accelerated free-breathing real-time myocardial tagging for exercise cardiovascular magnetic resonance.

Morales M, Yoon S, Fahmy A, Ghanbari F, Nakamori S, Rodriguez J J Cardiovasc Magn Reson. 2023; 25(1):56.

PMID: 37784153 PMC: 10544487. DOI: 10.1186/s12968-023-00961-w.


Multimodal phantoms for clinical PET/MRI.

Lennie E, Tsoumpas C, Sourbron S EJNMMI Phys. 2021; 8(1):62.

PMID: 34436671 PMC: 8390737. DOI: 10.1186/s40658-021-00408-0.


Synergistic motion compensation strategies for positron emission tomography when acquired simultaneously with magnetic resonance imaging.

Polycarpou I, Soultanidis G, Tsoumpas C Philos Trans A Math Phys Eng Sci. 2021; 379(2204):20200207.

PMID: 34218675 PMC: 8255946. DOI: 10.1098/rsta.2020.0207.


References
1.
Griswold M, Jakob P, Heidemann R, Nittka M, Jellus V, Wang J . Generalized autocalibrating partially parallel acquisitions (GRAPPA). Magn Reson Med. 2002; 47(6):1202-10. DOI: 10.1002/mrm.10171. View

2.
Deshmane A, Gulani V, Griswold M, Seiberlich N . Parallel MR imaging. J Magn Reson Imaging. 2012; 36(1):55-72. PMC: 4459721. DOI: 10.1002/jmri.23639. View

3.
Rahmim A, Dinelle K, Cheng J, Shilov M, Segars W, Lidstone S . Accurate event-driven motion compensation in high-resolution PET incorporating scattered and random events. IEEE Trans Med Imaging. 2008; 27(8):1018-33. PMC: 2920454. DOI: 10.1109/TMI.2008.917248. View

4.
Pruessmann K, Weiger M, Scheidegger M, Boesiger P . SENSE: sensitivity encoding for fast MRI. Magn Reson Med. 1999; 42(5):952-62. View

5.
Teras M, Kokki T, Durand-Schaefer N, Noponen T, Pietila M, Kiss J . Dual-gated cardiac PET-clinical feasibility study. Eur J Nucl Med Mol Imaging. 2009; 37(3):505-16. DOI: 10.1007/s00259-009-1252-4. View