In Vivo Evaluation of Edge-loading in Metal-on-metal Hip Resurfacing Patients with Pseudotumours
Overview
Authors
Affiliations
Objectives: Pseudotumours (abnormal peri-prosthetic soft-tissue reactions) following metal-on-metal hip resurfacing arthroplasty (MoMHRA) have been associated with elevated metal ion levels, suggesting that excessive wear may occur due to edge-loading of these MoM implants. This study aimed to quantify in vivo edge-loading in MoMHRA patients with and without pseudotumours during functional activities.
Methods: The duration and magnitude of edge-loading in vivo was quantified during functional activities by combining the dynamic hip joint segment contact force calculated from the three-dimensional (3D) motion analysis system with the 3D reconstruction of orientation of the acetabular component and each patient's specific hip joint centre, based on CT scans.
Results: Edge-loading in the hips with pseudotumours occurred with a four-fold increase in duration and magnitude of force compared with the hips without pseudotumours (p = 0.02).
Conclusions: The study provides the first in vivo evidence to support that edge-loading is an important mechanism that leads to localised excessive wear (edge-wear), with subsequent elevation of metal ion levels in MoMHRA patients with pseudotumours.
Capsular ligaments provide a passive stabilizing force to protect the hip against edge loading.
Karunaseelan K, Dandridge O, Muirhead-Allwood S, Van Arkel R, Jeffers J Bone Joint Res. 2021; 10(9):594-601.
PMID: 34555959 PMC: 8479567. DOI: 10.1302/2046-3758.109.BJR-2020-0536.R1.
Zheng N, Hu X, Dimitriou D, Dai K, Guo T, Tsai T Front Bioeng Biotechnol. 2021; 9:664907.
PMID: 34178961 PMC: 8222599. DOI: 10.3389/fbioe.2021.664907.
In vivo articular contact pattern of a total wrist arthroplasty design.
Akhbari B, Morton A, Shah K, Molino J, Moore D, Weiss A J Biomech. 2021; 121:110420.
PMID: 33895657 PMC: 8130596. DOI: 10.1016/j.jbiomech.2021.110420.
Li Z, Chen Z, Liu J, Liu F Orthop Surg. 2018; 10(4):312-320.
PMID: 30485684 PMC: 6594485. DOI: 10.1111/os.12408.
The impact of hip implant alignment on muscle and joint loading during dynamic activities.
Myers C, Laz P, Shelburne K, Judd D, Huff D, Winters J Clin Biomech (Bristol). 2018; 53:93-100.
PMID: 29482087 PMC: 5890933. DOI: 10.1016/j.clinbiomech.2018.02.010.