» Articles » PMID: 21520860

Validation for 2D/3D Registration. I: A New Gold Standard Data Set

Overview
Journal Med Phys
Specialty Biophysics
Date 2011 Apr 28
PMID 21520860
Citations 14
Authors
Affiliations
Soon will be listed here.
Abstract

Purpose: In this article, the authors propose a new gold standard data set for the validation of two-dimensional/three-dimensional (2D/3D) and 3D/3D image registration algorithms.

Methods: A gold standard data set was produced using a fresh cadaver pig head with attached fiducial markers. The authors used several imaging modalities common in diagnostic imaging or radiotherapy, which include 64-slice computed tomography (CT), magnetic resonance imaging using T1, T2, and proton density sequences, and cone beam CT imaging data. Radiographic data were acquired using kilovoltage and megavoltage imaging techniques. The image information reflects both anatomy and reliable fiducial marker information and improves over existing data sets by the level of anatomical detail, image data quality, and soft-tissue content. The markers on the 3D and 2D image data were segmented using ANALYZE 10.0 (AnalyzeDirect, Inc., Kansas City, KN) and an in-house software.

Results: The projection distance errors and the expected target registration errors over all the image data sets were found to be less than 2.71 and 1.88 mm, respectively.

Conclusions: The gold standard data set, obtained with state-of-the-art imaging technology, has the potential to improve the validation of 2D/3D and 3D/3D registration algorithms for image guided therapy.

Citing Articles

Application of frozen Thiel-embalmed specimens for radiotherapy delineation guideline development: a method to create accurate MRI-enhanced CT datasets.

Stouthandel M, Pullens P, Bogaert S, Schoepen M, Vangestel C, Achten E Strahlenther Onkol. 2022; 198(6):582-592.

PMID: 35403891 DOI: 10.1007/s00066-022-01928-z.


A new 2D-3D registration gold-standard dataset for the hip joint based on uncertainty modeling.

DIsidoro F, Chenes C, Ferguson S, Schmid J Med Phys. 2021; 48(10):5991-6006.

PMID: 34287934 PMC: 9290855. DOI: 10.1002/mp.15124.


Using synthetic CT for partial brain radiation therapy: Impact on image guidance.

Morris E, Price R, Kim J, Schultz L, Siddiqui M, Chetty I Pract Radiat Oncol. 2018; 8(5):342-350.

PMID: 29861348 PMC: 6123249. DOI: 10.1016/j.prro.2018.04.001.


Three-dimensional CBCT images registration method for TMJ based on reconstructed condyle and skull base.

Zhou Y, Li J, Lv W, Ma R, Li G Dentomaxillofac Radiol. 2018; 47(5):20170421.

PMID: 29595332 PMC: 6196037. DOI: 10.1259/dmfr.20170421.


A framework for automatic creation of gold-standard rigid 3D-2D registration datasets.

Madan H, Pernus F, Likar B, Spiclin Z Int J Comput Assist Radiol Surg. 2016; 12(2):263-275.

PMID: 27653616 DOI: 10.1007/s11548-016-1482-4.


References
1.
Fitzpatrick J, West J, Maurer Jr C . Predicting error in rigid-body point-based registration. IEEE Trans Med Imaging. 1999; 17(5):694-702. DOI: 10.1109/42.736021. View

2.
Xing L, Thorndyke B, Schreibmann E, Yang Y, Li T, Kim G . Overview of image-guided radiation therapy. Med Dosim. 2006; 31(2):91-112. DOI: 10.1016/j.meddos.2005.12.004. View

3.
Groh B, Siewerdsen J, Drake D, Wong J, Jaffray D . A performance comparison of flat-panel imager-based MV and kV cone-beam CT. Med Phys. 2002; 29(6):967-75. DOI: 10.1118/1.1477234. View

4.
van de Kraats E, Penney G, Tomazevic D, van Walsum T, Niessen W . Standardized evaluation methodology for 2-D-3-D registration. IEEE Trans Med Imaging. 2005; 24(9):1177-89. DOI: 10.1109/TMI.2005.853240. View

5.
Wang M, Maurer Jr C, Fitzpatrick J, Maciunas R . An automatic technique for finding and localizing externally attached markers in CT and MR volume images of the head. IEEE Trans Biomed Eng. 1996; 43(6):627-37. DOI: 10.1109/10.495282. View