Fiducial Point Placement and the Accuracy of Point-based, Rigid Body Registration

Overview
Journal Neurosurgery
Specialty Neurosurgery
Date 2001 Apr 27
PMID 11322441
Citations 75
Authors
Affiliations
Soon will be listed here.
Abstract

Objective: To demonstrate that the shape of the configuration of fiducial points is an important factor governing target registration error (TRE) in point-based, rigid registration.

Methods: We consider two clinical situations: cranial neurosurgery and pedicle screw placement. For cranial neurosurgery, we apply theoretical results concerning TRE prediction, which we have previously derived and validated, to three hypothetical fiducial marker configurations. We illustrate the profile of expected TRE for each configuration. For pedicle screw placement, we apply the same theory to a common anatomic landmark configuration (tips of spinous and transverse processes) used for pedicle screw placement, and we estimate the error rate expected in placement of the screw.

Results: In the cranial neurosurgery example, we demonstrate that relatively small values of TRE may be achieved by using widely spread fiducial markers and/or placing the centroid of the markers near the target. We also demonstrate that near-collinear marker configurations far from the target may result in large TRE values. In the pedicle screw placement example, we demonstrate that the screw must be approximately 4 mm narrower than the pedicle in which it is implanted to minimize the chance of pedicle violation during placement.

Conclusion: The placement of fiducial points is an important factor in minimizing the error rate for point-based, rigid registration. By using as many points as possible, avoiding near-collinear configurations, and ensuring that the centroid of the fiducial points is as near as possible to the target, TREs can be minimized.

Citing Articles

Augmented Reality in Extratemporal Lobe Epilepsy Surgery.

Grote A, Neumann F, Menzler K, Carl B, Nimsky C, Bopp M J Clin Med. 2024; 13(19).

PMID: 39407752 PMC: 11477171. DOI: 10.3390/jcm13195692.


Noctopus: a novel device and method for patient registration and navigation in image-guided cranial surgery.

Ozbek Y, Bardosi Z, Freysinger W Int J Comput Assist Radiol Surg. 2024; 19(12):2371-2380.

PMID: 38748051 PMC: 11607009. DOI: 10.1007/s11548-024-03135-w.


Generating Patient-Specific Acoustic Simulations for Transcranial Focused Ultrasound Procedures Based on Optical Tracking Information.

Sigona M, Manuel T, Phipps M, Banaie Boroujeni K, Treuting R, Womelsdorf T IEEE Open J Ultrason Ferroelectr Freq Control. 2024; 3:146-156.

PMID: 38222464 PMC: 10785958. DOI: 10.1109/ojuffc.2023.3318560.


The Role of Augmented Reality Neuronavigation in Transsphenoidal Surgery: A Systematic Review.

Campisi B, Costanzo R, Gulino V, Avallone C, Noto M, Bonosi L Brain Sci. 2023; 13(12).

PMID: 38137143 PMC: 10741598. DOI: 10.3390/brainsci13121695.


Influence of surgical position and registration methods on clinical accuracy of navigation systems in brain tumor surgery.

Furuse M, Ikeda N, Kawabata S, Park Y, Takeuchi K, Fukumura M Sci Rep. 2023; 13(1):2644.

PMID: 36788314 PMC: 9929322. DOI: 10.1038/s41598-023-29710-w.