» Articles » PMID: 19582531

A Multicenter Evaluation of a New Post-processing Method with Depth-dependent Collimator Resolution Applied to Full-time and Half-time Acquisitions Without and with Simultaneously Acquired Attenuation Correction

Overview
Journal J Nucl Cardiol
Date 2009 Jul 8
PMID 19582531
Citations 45
Authors
Affiliations
Soon will be listed here.
Abstract

Background: The field of nuclear cardiology is limited by image quality and length of procedure. The use of depth-dependent resolution recovery algorithms in conjunction with iterative reconstruction holds promise to improve image quality and reduce acquisition time. This study compared the Astonish algorithm employing depth-dependent resolution recovery and iterative reconstruction to filtered backprojection (FBP) using both full-time (FTA) and half-time (HTA) data. Attenuation correction including scatter correction in conjunction with the Astonish algorithm was also evaluated.

Methods: We studied 187 consecutive patients (132 with cardiac catheterization and 55 with low likelihood for CAD) from three nuclear cardiology laboratories who had previously undergone clinically indicated rest/stress Tc-99m sestamibi or tetrofosmin SPECT. Acquisition followed ASNC guidelines (64 projections, 20-25 seconds). Processing of the full-time data sets included FBP and Astonish (FTA). A total of 32 projection data sets were created by stripping the full-time data sets and processing with Astonish (HTA). Attenuation correction was applied to both full-time and half-time Astonish-processed images (FTA-AC and HTA-AC, respectively). A consensus interpretation of three blinded readers was performed for image quality, interpretative certainty, and diagnostic accuracy, as well as severity and reversibility of perfusion and functional parameters.

Results: Full-time and half-time Astonish processing resulted in a significant improvement in image quality in comparison with FBP. Stress and rest perfusion image quality (excellent or good) were 85%/80% (FBP), 98%/95% (FTA), and 95%/92% (HTA), respectively (p < 0.001). Interpretative certainty and diagnostic accuracy were similar with FBP, FTA, and HTA. Left ventricular functional data were not different despite a slight reduction in half-time gated image quality. Application of attenuation correction resulted in similar image quality and improved normalcy (FTA vs. FTA-AC: 76% vs. 95%; HTA vs. HTA-AC: 76% vs. 100%) and specificity (FTA vs. FTA-AC: 62% vs. 78%; HTA vs. HTA-AC: 63% vs. 84%) (p < 0.01 for all comparisons).

Conclusion: Astonish processing, which incorporates depth-dependent resolution recovery, improves image quality without sacrificing interpretative certainty or diagnostic accuracy. Application of simultaneously acquired attenuation correction, which includes scatter correction, to full-time and half-time images processed with this method, improves specificity and normalcy while maintaining high image quality.

Citing Articles

Clinical Utility of CT-based Attenuation-correction in Myocardial Perfusion SPECT Imaging.

Hatipoglu F, Cetin N Mol Imaging Radionucl Ther. 2023; 32(2):138-145.

PMID: 37337825 PMC: 10284174. DOI: 10.4274/mirt.galenos.2022.68094.


Simulating dose reduction for myocardial perfusion SPECT using a Poisson resampling method.

Kim I, Lee S, An Y, Choi S, Yoon J Nucl Med Mol Imaging. 2021; 55(5):245-252.

PMID: 34721717 PMC: 8517053. DOI: 10.1007/s13139-021-00710-w.


Drivers of radiation dose reduction with myocardial perfusion imaging: A large health system experience.

Al Badarin F, Spertus J, Bateman T, Patel K, Burgett E, Kennedy K J Nucl Cardiol. 2019; 27(3):785-794.

PMID: 30706351 PMC: 6669103. DOI: 10.1007/s12350-018-01576-w.


Contemporary Cardiac SPECT Imaging-Innovations and Best Practices: An Information Statement from the American Society of Nuclear Cardiology.

Abbott B, Case J, Dorbala S, Einstein A, Galt J, Pagnanelli R J Nucl Cardiol. 2018; 25(5):1847-1860.

PMID: 30143954 DOI: 10.1007/s12350-018-1348-y.


Myocardial perfusion imaging in women for the evaluation of stable ischemic heart disease-state-of-the-evidence and clinical recommendations.

Taqueti V, Dorbala S, Wolinsky D, Abbott B, Heller G, Bateman T J Nucl Cardiol. 2017; 24(4):1402-1426.

PMID: 28585034 PMC: 5942593. DOI: 10.1007/s12350-017-0926-8.


References
1.
Bland J, Altman D . Statistical methods for assessing agreement between two methods of clinical measurement. Lancet. 1986; 1(8476):307-10. View

2.
Henzlova M, Cerqueira M, Mahmarian J, Yao S . Stress protocols and tracers. J Nucl Cardiol. 2006; 13(6):e80-90. DOI: 10.1016/j.nuclcard.2006.08.011. View

3.
Noble G, Ahlberg A, Kokkirala A, Cullom S, Bateman T, Cyr G . Validation of attenuation correction using transmission truncation compensation with a small field of view dedicated cardiac SPECT camera system. J Nucl Cardiol. 2009; 16(2):222-32. DOI: 10.1007/s12350-008-9022-4. View

4.
Thompson R, Heller G, Johnson L, Case J, Cullom S, Garcia E . Value of attenuation correction on ECG-gated SPECT myocardial perfusion imaging related to body mass index. J Nucl Cardiol. 2005; 12(2):195-202. DOI: 10.1016/j.nuclcard.2004.12.298. View

5.
Case J, Hsu B, Bateman T, Cullom S . A Bayesian iterative transmission gradient reconstruction algorithm for cardiac SPECT attenuation correction. J Nucl Cardiol. 2007; 14(3):324-33. DOI: 10.1016/j.nuclcard.2007.02.004. View