First Experience with 320-row Multidetector CT Coronary Angiography Scanning with Prospective Electrocardiogram Gating to Reduce Radiation Dose

Overview
Publisher Elsevier
Date 2009 Jul 7
PMID 19577215
Citations 20
Authors
Affiliations
Soon will be listed here.
Abstract

Background: The 320-row multidetector CT (MDCT) provides the capability for prospective electrocardiogram-gated cardiac CT angiography.

Objective: We assessed the effective radiation doses of MDCT with a 320-row detector volume scanner.

Methods: Two hundred patients underwent clinically directed cardiac scanning (100 kVp, n=9 or 120 kVp, n=191; range, 300-580 mA). Effective radiation dose (in mSv) was estimated from extended dose-length product. For heart rates (HRs) <65 beats/min, exposure phase was 65% up to the end of R wave of the cardiac cycle, using a one-heart beat acquisition and half-scan reconstruction. HRs from 66 to 79 beats/min and > or =80 beats/min were scanned with either 2- or 3-heart beat acquisitions, respectively. Patients with arrhythmias were not excluded.

Results: The mean effective dose was 5.7+/-1.7 mSv (range, 1.6-11.1 mSv) for 151 patients (75%) scanned with a one-heart beat acquisition. Qualitative image quality was assessed to be in good to excellent range, and mottle image quality was in low-to-medium mottle range. For patients scanned with 2- or 3-heart beat acquisition, radiation dose was higher with mean exposures of 13.0+/-3.3 mSv and 19.5+/-5.3 mSv, respectively.

Conclusions: Low effective radiation dose with acceptable image quality on 320-row MDCT can be achieved with one-heart beat scan acquisition when HR is <65 beats/min. Further reduction in dose can likely also be achieved by modification of the prospective-gated imaging parameters.

Citing Articles

Advances in physical diagnosis and treatment of male erectile dysfunction.

Pang K, Pan D, Xu H, Ma Y, Wang J, Xu P Front Physiol. 2023; 13:1096741.

PMID: 36699684 PMC: 9868413. DOI: 10.3389/fphys.2022.1096741.


Diagnosis of coronary artery disease in patients with atrial fibrillation using low tube voltage coronary CT angiography with isotonic low-concentration contrast agent.

Pan Y, Huang Q, Zhu Y, Zou X, Chu H, Du X Int J Cardiovasc Imaging. 2019; 35(12):2239-2248.

PMID: 31363878 DOI: 10.1007/s10554-019-01678-7.


Submillisievert imaging protocol using full reconstruction and advanced patient motion correction in 320-row area detector coronary CT angiography.

Kawaguchi Y, Fujimoto S, Takamura K, Kato E, Suda S, Matsumori R Int J Cardiovasc Imaging. 2017; 34(3):465-474.

PMID: 28900781 DOI: 10.1007/s10554-017-1237-5.


Application of low-dose dual-source computed tomography angiography in children with complex congenital heart disease.

Chen X, Jiang F, Li L, Chen Y, Chen X, Jiang Y Exp Ther Med. 2017; 14(2):1177-1183.

PMID: 28810576 PMC: 5526138. DOI: 10.3892/etm.2017.4591.


Role of 256-slice MDCT in the evaluation of coronary artery fistula: A case series with review of literature.

Dashottar S, Singh R, Malani S, Sarin A, Arora H Med J Armed Forces India. 2016; 72(4):393-399.

PMID: 27843191 PMC: 5099439. DOI: 10.1016/j.mjafi.2016.06.001.