» Articles » PMID: 12930558

Do Changes in Pulse Oximeter Oxygen Saturation Predict Equivalent Changes in Arterial Oxygen Saturation?

Overview
Journal Crit Care
Specialty Critical Care
Date 2003 Aug 22
PMID 12930558
Citations 54
Authors
Affiliations
Soon will be listed here.
Abstract

Introduction: This study investigates the relation between changes in pulse oximeter oxygen saturation (SpO2) and changes in arterial oxygen saturation (SaO2) in the critically ill, and the effects of acidosis and anaemia on precision of using pulse oximetry to predict SaO2.

Patients And Methods: Forty-one consecutive patients were recruited from a nine-bed general intensive care unit into a 2-month study. Patients with significant jaundice (bilirubin >40 micromol/l) or inadequate pulse oximetry tracing were excluded.

Results: A total of 1085 paired readings demonstrated only moderate correlation (r= 0.606; P < 0.01) between changes in SpO2 and those in SaO2, and the pulse oximeter tended to overestimate actual changes in SaO2. Anaemia increased the degree of positive bias whereas acidosis reduced it. However, the magnitude of these changes was small.

Conclusion: Changes in SpO2 do not reliably predict equivalent changes in SaO2 in the critically ill. Neither anaemia nor acidosis alters the relation between SpO2 and SaO2 to any clinically important extent.

Citing Articles

Limitations of SpO / FiO-ratio for classification and monitoring of acute respiratory distress syndrome-an observational cohort study.

Erlebach R, Pale U, Beck T, Markovic S, Seric M, David S Crit Care. 2025; 29(1):82.

PMID: 39972458 PMC: 11837723. DOI: 10.1186/s13054-025-05317-7.


Assessment of Initial Oxygenation Levels of Chronic Obstructive Pulmonary Disease (COPD) and Their Impact on Basis and Vital Tools: Retrospective Cohort Study From India.

Ghosh S, Brahmachari R, Ghosh S, Das Choudhury S, Ghosh K Cureus. 2025; 16(12):e75470.

PMID: 39791024 PMC: 11717386. DOI: 10.7759/cureus.75470.


Utility of Skin Tone on Pulse Oximetry in Critically Ill Patients: A Prospective Cohort Study.

Hao S, Dempsey K, Matos J, Cox C, Rotemberg V, Gichoya J Crit Care Explor. 2024; 6(9):e1133.

PMID: 39268149 PMC: 11392475. DOI: 10.1097/CCE.0000000000001133.


Utility of skin tone on pulse oximetry in critically ill patients: a prospective cohort study.

Hao S, Dempsey K, Matos J, Cox C, Rotemberg V, Gichoya J medRxiv. 2024; .

PMID: 38464170 PMC: 10925348. DOI: 10.1101/2024.02.24.24303291.


Accuracy of the Apple Watch in measuring oxygen saturation: comparison with pulse oximetry and ABG.

Arslan B, Sener K, Guven R, Kapci M, Korkut S, Sutasir M Ir J Med Sci. 2023; 193(1):477-483.

PMID: 37440093 DOI: 10.1007/s11845-023-03456-w.


References
1.
Jubran A . Advances in respiratory monitoring during mechanical ventilation. Chest. 1999; 116(5):1416-25. DOI: 10.1378/chest.116.5.1416. View

2.
Ortiz F, Aldrich T, Nagel R, Benjamin L . Accuracy of pulse oximetry in sickle cell disease. Am J Respir Crit Care Med. 1999; 159(2):447-51. DOI: 10.1164/ajrccm.159.2.9806108. View

3.
Van de Louw A, Cracco C, Cerf C, Harf A, Duvaldestin P, Lemaire F . Accuracy of pulse oximetry in the intensive care unit. Intensive Care Med. 2001; 27(10):1606-13. DOI: 10.1007/s001340101064. View

4.
Bland J, Altman D . Statistical methods for assessing agreement between two methods of clinical measurement. Lancet. 1986; 1(8476):307-10. View

5.
Turner J, McNicol M, Sillett R . Distribution of carboxyhaemoglobin concentrations in smokers and non-smokers. Thorax. 1986; 41(1):25-7. PMC: 460247. DOI: 10.1136/thx.41.1.25. View