A Review of the Effect of Skin Pigmentation on Pulse Oximeter Accuracy

Overview
Journal Physiol Meas
Date 2023 May 12
PMID 37172609
Authors
Affiliations
Soon will be listed here.
Abstract

. Pulse oximetry is a non-invasive optical technique used to measure arterial oxygen saturation (SpO) in a variety of clinical settings and scenarios. Despite being one the most significant technological advances in health monitoring over the last few decades, there have been reports on its various limitations. Recently due to the Covid-19 pandemic, questions about pulse oximeter technology and its accuracy when used in people with different skin pigmentation have resurfaced, and are to be addressed.. This review presents an introduction to the technique of pulse oximetry including its basic principle of operation, technology, and limitations, with a more in depth focus on skin pigmentation. Relevant literature relating to the performance and accuracy of pulse oximeters in populations with different skin pigmentation are evaluated.. The majority of the evidence suggests that the accuracy of pulse oximetry differs in subjects of different skin pigmentations to a level that requires particular attention, with decreased accuracy in patients with dark skin.. Some recommendations, both from the literature and contributions from the authors, suggest how future work could address these inaccuracies to potentially improve clinical outcomes. These include the objective quantification of skin pigmentation to replace currently used qualitative methods, and computational modelling for predicting calibration algorithms based on skin colour.

Citing Articles

Modality-level obstacles and initiatives to improve representation in fetal, infant, and toddler neuroimaging research samples.

Margolis E, Nelson P, Fiske A, Champaud J, Olson H, Gomez M Dev Cogn Neurosci. 2025; 72:101505.

PMID: 39954600 PMC: 11875194. DOI: 10.1016/j.dcn.2024.101505.


Monte Carlo Simulation of the Effect of Melanin Concentration on Light-Tissue Interactions in Reflectance Pulse Oximetry.

Al-Halawani R, Qassem M, Kyriacou P Sensors (Basel). 2025; 25(2).

PMID: 39860931 PMC: 11769186. DOI: 10.3390/s25020559.


Analysis of the Effect of Skin Pigmentation and Oxygen Saturation on Monte Carlo-Simulated Reflectance Photoplethysmography Signals.

Al-Halawani R, Qassem M, Kyriacou P Sensors (Basel). 2025; 25(2.

PMID: 39860743 PMC: 11769505. DOI: 10.3390/s25020372.


Light source spectra and higher measurement uncertainty in pulse oximeter readings for individuals with darker skin pigmentation.

Benner K, Bierman A, Rea M Br J Anaesth. 2025; 134(3):856-858.

PMID: 39818456 PMC: 11867059. DOI: 10.1016/j.bja.2024.11.034.


Development and characterization of silicone-based tissue phantoms for pulse oximeter performance testing.

Bhusal A, Farahmand M, Hasan M, Vasudevan S, Vogt W, Ibarra B J Biomed Opt. 2025; 29(Suppl 3):S33314.

PMID: 39776836 PMC: 11706025. DOI: 10.1117/1.JBO.29.S3.S33314.


References
1.
Ries A, Prewitt L, Johnson J . Skin color and ear oximetry. Chest. 1989; 96(2):287-90. DOI: 10.1378/chest.96.2.287. View

2.
Teo J . Early Detection of Silent Hypoxia in Covid-19 Pneumonia Using Smartphone Pulse Oximetry. J Med Syst. 2020; 44(8):134. PMC: 7305055. DOI: 10.1007/s10916-020-01587-6. View

3.
Bangash M, Hodson J, Evison F, Patel J, Johnston A, Gallier S . Impact of ethnicity on the accuracy of measurements of oxygen saturations: A retrospective observational cohort study. EClinicalMedicine. 2022; 48:101428. PMC: 9096912. DOI: 10.1016/j.eclinm.2022.101428. View

4.
Tytler J, Seeley H . The Nellcor N-101 pulse oximeter. A clinical evaluation in anaesthesia and intensive care. Anaesthesia. 1986; 41(3):302-5. DOI: 10.1111/j.1365-2044.1986.tb12793.x. View

5.
Jubran A . Pulse oximetry. Crit Care. 2015; 19:272. PMC: 4504215. DOI: 10.1186/s13054-015-0984-8. View