Evaluation of Cortical Bone by Computed Tomography

Overview
Date 1996 Oct 1
PMID 8889852
Citations 52
Authors
Affiliations
Soon will be listed here.
Abstract

The purpose of this study was to determine the minimum thickness of cortical bone required for the accurate measurement of cortical material density by computed tomography (CT) and to establish normal reference values. A phantom with several wall thicknesses of bone-like material was constructed to simulate various cortical widths. The CT density at each level of thickness was measured on a GE 9800 CT scanner and on the OsteoQuant, a special CT scanner optimized for the measurement of bone in the extremities. The minimum width required to attain the correct material density was determined for each scanner. Additionally, the material density and width of the cortex in the radius and/or femur were measured by CT in 761 healthy subjects, ages 4-84 years. The minimum thickness necessary for an accurate density evaluation of the walls of the phantom by CT was 2-2.5 mm; below these thresholds the values fell in a linear way relative to width. In humans, the material density of cortical bone in the appendicular skeleton was not influenced by height or weight, and the values were similar for all subjects, as long as the cortical width was above 2-2.5 mm. The cortical width increased with age up to 30 years and decreased from 50 years on. We conclude that the material density of cortical bone in the appendicular skeleton can be measured accurately by CT if the thickness of the cortex exceeds 2-2.5 mm.

Citing Articles

Measuring the thickness of vertebral endplate and shell using digital tomosynthesis.

Yeni Y, Dix M, Xiao A, Oravec D, Flynn M Bone. 2022; 157:116341.

PMID: 35092890 PMC: 8858866. DOI: 10.1016/j.bone.2022.116341.


Biological basis of bone strength: anatomy, physiology and measurement.

Hart N, Newton R, Tan J, Rantalainen T, Chivers P, Siafarikas A J Musculoskelet Neuronal Interact. 2020; 20(3):347-371.

PMID: 32877972 PMC: 7493450.


Osteoarthritis Changes Hip Geometry and Biomechanics Regardless of Bone Mineral Density-A Quantitative Computed Tomography Study.

Narloch J, Glinkowski W J Clin Med. 2019; 8(5).

PMID: 31083644 PMC: 6572464. DOI: 10.3390/jcm8050669.


Fracture Prediction by Computed Tomography and Finite Element Analysis: Current and Future Perspectives.

Johannesdottir F, Allaire B, Bouxsein M Curr Osteoporos Rep. 2018; 16(4):411-422.

PMID: 29846870 DOI: 10.1007/s11914-018-0450-z.


Meagre effects of disuse on the human fibula are not explained by bone size or geometry.

Ireland A, Capozza R, Cointry G, Nocciolino L, Ferretti J, Rittweger J Osteoporos Int. 2016; 28(2):633-641.

PMID: 27734100 DOI: 10.1007/s00198-016-3779-0.