Placement of a Ferromagnetic Intracerebral Aneurysm Clip in a Magnetic Field with a Fatal Outcome

Overview
Journal Radiology
Specialty Radiology
Date 1993 Jun 1
PMID 8497645
Citations 21
Authors
Affiliations
Soon will be listed here.
Abstract

Magnetic resonance (MR) imaging may be contraindicated in patients with biomedical devices, among the most dangerous of which are intracranial aneurysm clips, owing to the possibility of torque and dislodgement. A case is presented in which a patient with a reportedly nonferromagnetic clip was placed in a magnetic field. The patient developed an acute intracerebral hemorrhage in the MR unit, with a fatal outcome. Imaging studies strongly suggested a torqued clip as the cause. Autopsy revealed a torn middle cerebral artery from clip movement, and the clip was identified as a ferromagnetic type. This is the first reported case, to the authors' knowledge, of a fatal outcome due to an intracranial aneurysm clip placed in a magnetic field.

Citing Articles

A framework for developing generic implant safety procedures for scanning patients with medical implants and devices in MRI.

Ashmore J, Prescott S, McLean J, Wilson D, Charles-Edwards G, Wright P Br J Radiol. 2024; 98(1167):336-344.

PMID: 39535863 PMC: 11840167. DOI: 10.1093/bjr/tqae232.


MRI following scoliosis surgery? An analysis of implant heating, displacement, torque, and susceptibility artifacts.

Heinrich A, Reinhold M, Guttler F, Matziolis G, Teichgraber U, Zippelius T Eur Radiol. 2020; 31(6):4298-4307.

PMID: 33277671 PMC: 8128815. DOI: 10.1007/s00330-020-07546-6.


The Zurich Checklist for Safety in the Intraoperative Magnetic Resonance Imaging Suite: Technical Note.

Stienen M, Fierstra J, Pangalu A, Regli L, Bozinov O Oper Neurosurg (Hagerstown). 2018; 16(6):756-765.

PMID: 30099512 PMC: 6531895. DOI: 10.1093/ons/opy205.


[MR implant labelling and its use in clinical MRI practice].

Muhlenweg M, Schaefers G Radiologe. 2015; 55(8):682-90.

PMID: 26296804 DOI: 10.1007/s00117-015-2814-z.


[Whole-body ferromagnetic detector systems in clinical MRI].

Guttler F, Heinrich A, Teichgraber U Radiologe. 2015; 55(8):649-53.

PMID: 26245984 DOI: 10.1007/s00117-015-2813-0.