» Articles » PMID: 37875769

Photon-counting Detector CT-based Virtual Monoenergetic Reconstructions: Repeatability and Reproducibility of Radiomics Features of an Organic Phantom and Human Myocardium

Abstract

Background: Photon-counting detector computed tomography (PCD-CT) may influence imaging characteristics for various clinical conditions due to higher signal and contrast-to-noise ratio in virtual monoenergetic images (VMI). Radiomics analysis relies on quantification of image characteristics. We evaluated the impact of different VMI reconstructions on radiomic features in in vitro and in vivo PCD-CT datasets.

Methods: An organic phantom consisting of twelve samples (four oranges, four onions, and four apples) was scanned five times. Twenty-three patients who had undergone coronary computed tomography angiography on a first generation PCD-CT system with the same image acquisitions were analyzed. VMIs were reconstructed at 6 keV levels (40, 55, 70, 90, 120, and 190 keV). The phantoms and the patients' left ventricular myocardium (LVM) were segmented for all reconstructions. Ninety-three original radiomic features were extracted. Repeatability and reproducibility were evaluated through intraclass correlations coefficient (ICC) and post hoc paired samples ANOVA t test.

Results: There was excellent repeatability for radiomic features in phantom scans (all ICC = 1.00). Among all VMIs, 36/93 radiomic features (38.7%) in apples, 28/93 (30.1%) in oranges, and 33/93 (35.5%) in onions were not significantly different. For LVM, the percentage of stable features was high between VMIs ≥ 90 keV (90 versus 120 keV, 77.4%; 90 versus 190 keV, 83.9%; 120 versus 190 keV, 89.3%), while comparison to lower VMI levels led to fewer reproducible features (40 versus 55 keV, 8.6%).

Conclusions: VMI levels influence the stability of radiomic features in an organic phantom and patients' LVM; stability decreases considerably below 90 keV.

Relevance Statement: Spectral reconstructions significantly influence radiomic features in vitro and in vivo, necessitating standardization and careful attention to these reconstruction parameters before clinical implementation.

Key Points: • Radiomic features have an excellent repeatability within the same PCD-CT acquisition and reconstruction. • Differences in VMI lead to decreased reproducibility for radiomic features. • VMI ≥ 90 keV increased the reproducibility of the radiomic features.

Citing Articles

Value of Ultrahigh-Resolution Photon-Counting Detector Computed Tomography in Cardiac Imaging.

Kravchenko D, Hagar M, Vecsey-Nagy M, Tremamunno G, Szilveszter B, Vattay B Echocardiography. 2025; 42(2):e70100.

PMID: 39945196 PMC: 11822751. DOI: 10.1111/echo.70100.


Robustness of radiomics within photon-counting detector CT: impact of acquisition and reconstruction factors.

Zhang H, Lu T, Wang L, Xing Y, Hu Y, Xu Z Eur Radiol. 2025; .

PMID: 39890616 DOI: 10.1007/s00330-025-11374-x.


Radiomics Feature Stability in True and Virtual Non-Contrast Reconstructions from Cardiac Photon-Counting Detector CT Datasets.

Canalini L, Becker E, Risch F, Bette S, Hellbrueck S, Becker J Diagnostics (Basel). 2024; 14(22).

PMID: 39594149 PMC: 11592515. DOI: 10.3390/diagnostics14222483.


Intraindividual reproducibility of myocardial radiomic features between energy-integrating detector and photon-counting detector CT angiography.

Tremamunno G, Varga-Szemes A, Schoepf U, Laghi A, Zsarnoczay E, Fink N Eur Radiol Exp. 2024; 8(1):101.

PMID: 39196286 PMC: 11358367. DOI: 10.1186/s41747-024-00493-7.


Robustness of radiomics among photon-counting detector CT and dual-energy CT systems: a texture phantom study.

Zhu L, Dong H, Sun J, Wang L, Xing Y, Hu Y Eur Radiol. 2024; 35(2):871-884.

PMID: 39048741 PMC: 11782343. DOI: 10.1007/s00330-024-10976-1.


References
1.
Khan J, Singh A, Nazir S, Kanagala P, Gershlick A, McCann G . Comparison of cardiovascular magnetic resonance feature tracking and tagging for the assessment of left ventricular systolic strain in acute myocardial infarction. Eur J Radiol. 2015; 84(5):840-8. DOI: 10.1016/j.ejrad.2015.02.002. View

2.
Gillies R, Kinahan P, Hricak H . Radiomics: Images Are More than Pictures, They Are Data. Radiology. 2015; 278(2):563-77. PMC: 4734157. DOI: 10.1148/radiol.2015151169. View

3.
Park B, Kim J, Heo C, Park K . Reliability of CT radiomic features reflecting tumour heterogeneity according to image quality and image processing parameters. Sci Rep. 2020; 10(1):3852. PMC: 7052198. DOI: 10.1038/s41598-020-60868-9. View

4.
Tian X, Sun C, Liu Z, Li W, Duan H, Wang L . Prediction of Response to Preoperative Neoadjuvant Chemotherapy in Locally Advanced Cervical Cancer Using Multicenter CT-Based Radiomic Analysis. Front Oncol. 2020; 10:77. PMC: 7010718. DOI: 10.3389/fonc.2020.00077. View

5.
Huang Y, Liang C, He L, Tian J, Liang C, Chen X . Development and Validation of a Radiomics Nomogram for Preoperative Prediction of Lymph Node Metastasis in Colorectal Cancer. J Clin Oncol. 2016; 34(18):2157-64. DOI: 10.1200/JCO.2015.65.9128. View