Experimental Investigation of the Effective Point of Measurement for Plane-parallel Chambers Used in Electron Beam Dosimetry

Overview
Date 2023 Jun 12
PMID 37307247
Authors
Affiliations
Soon will be listed here.
Abstract

In this study, the effective point of measurement (EPOM) for plane-parallel ionization chambers in clinical high-energy electron beams was determined experimentally. Previous studies have reported that the EPOM of plane-parallel chambers is shifted several tens of millimeters downstream from the inner surface of the entrance window to the cavity. These findings were based on the Monte Carlo (MC) simulation, and few experimental studies have been performed. Thus, additional experimental validations of the reported EPOMs were required. In this study, we investigated the EPOMs of three plane-parallel chambers (NACP-02, Roos and Advanced Markus) for clinical electron beams. The EPOMs were determined by comparing the measured percentage depth-dose (PDD) of the plane-parallel chambers and the PDD obtained using the microDiamond detector. The optimal shift to the EPOM was energy-dependent. The determined EPOM showed no chamber-to-chamber variation, thereby allowing the use of a single value. The mean optimal shifts were 0.104 ± 0.011, 0.040 ± 0.012, and 0.012 ± 0.009 cm for NACP-02, Roos, and Advanced Markus, respectively. These values are valid in the R range from 2.40 to 8.82 cm, which correspond to 6-22 MeV. Roos and Advanced Markus exhibited similar results to those of the previous studies, but NACP-02 showed a larger shift. This is probably due to the uncertainty of the entrance window of NACP-02. Therefore, it is necessary to carefully consider where the optimal EPOM is located when using this chamber.

Citing Articles

Experimental investigation of the effective point of measurement for plane-parallel chambers used in electron beam dosimetry.

Yasui K, Nakajima Y, Suda Y, Arai Y, Takizawa T, Sakai K J Appl Clin Med Phys. 2023; 24(7):e14059.

PMID: 37307247 PMC: 10338742. DOI: 10.1002/acm2.14059.

References
1.
Muir B, McEwen M, Rogers D . Determination of relative ion chamber calibration coefficients from depth-ionization measurements in clinical electron beams. Phys Med Biol. 2014; 59(19):5953-69. DOI: 10.1088/0031-9155/59/19/5953. View

2.
Pawiro S, Mahfirotin D, Assegab M, Wibowo W . Modified electron beam output calibration based on IAEA Technical Report Series 398. J Appl Clin Med Phys. 2022; 23(4):e13573. PMC: 8992941. DOI: 10.1002/acm2.13573. View

3.
McEwen M, DuSautoy A, Williams A . The calibration of therapy level electron beam ionization chambers in terms of absorbed dose to water. Phys Med Biol. 1998; 43(9):2503-19. DOI: 10.1088/0031-9155/43/9/006. View

4.
Araki F . Monte Carlo calculations of correction factors for plane-parallel ionization chambers in clinical electron dosimetry. Med Phys. 2008; 35(9):4033-40. DOI: 10.1118/1.2968102. View

5.
Chin E, Shipley D, Bailey M, Seuntjens J, Palmans H, Dusautoy A . Validation of a Monte Carlo model of a NACP-02 plane-parallel ionization chamber model using electron backscatter experiments. Phys Med Biol. 2008; 53(8):N119-26. DOI: 10.1088/0031-9155/53/8/N01. View