Automatic Measurement of Beam-positioning Accuracy at Off-isocenter Positions

Overview
Date 2022 Nov 24
PMID 36420973
Authors
Affiliations
Soon will be listed here.
Abstract

Purpose: This study performed an automatic measurement of the off-axis beam-positioning accuracy at a single isocenter via the TrueBeam Developer mode and evaluated the beam-positioning accuracy considering the effect of couch rotational errors.

Methods: TrueBeam STx and the Winston-Lutz test-dedicated phantom, with a 3 mm diameter steel ball, were used in this study. The phantom was placed on the treatment couch, and the Winston-Lutz test was performed at the isocenter for four gantry angles (0°, 90°, 180°, and 270°) using an electronic portal imaging device. The phantom offset positions were at distances of 0, 25, 50, 75, and 100 mm from the isocenter along the superior-inferior, anterior-posterior, and left-right directions. Seventeen patterns of multileaf collimator-shaped square fields of 10 × 10 mm were created at the isocenter and off-axis positions for each gantry angle. The beam-positioning accuracy was evaluated with couch rotation along the yaw-axis (0°, ± 0.5°, and ± 1.0°).

Results: The mean beam-positioning errors at the isocenter and off-isocenter distances (from the isocenter to ±100 mm) were 0.46-0.60, 0.44-0.91, and 0.42-1.11 mm for the couch angles of 0°, ±0.5°, and ±1°, respectively. The beam-positioning errors increased as the distance from the isocenter and couch rotation increased.

Conclusion: These findings suggest that the beam-positioning accuracy at the isocenter and off-isocenter positions can be evaluated quickly and automatically using the TrueBeam Developer mode. The proposed procedure is expected to contribute to an efficient evaluation of the beam-positioning accuracy at off-isocenter positions.

Citing Articles

Automatic measurement of beam-positioning accuracy at off-isocenter positions.

Ono T, Kido T, Nakamura M, Iramina H, Kakino R, Mizowaki T J Appl Clin Med Phys. 2022; 24(3):e13844.

PMID: 36420973 PMC: 10018661. DOI: 10.1002/acm2.13844.

References
1.
Wack L, Exner F, Wegener S, Sauer O . The impact of isocentric shifts on delivery accuracy during the irradiation of small cerebral targets-Quantification and possible corrections. J Appl Clin Med Phys. 2020; 21(5):56-64. PMC: 7286018. DOI: 10.1002/acm2.12854. View

2.
Rose M, Tirpak L, Van Casteren K, Zack J, Simon T, Schoenfeld A . Multi-institution validation of a new high spatial resolution diode array for SRS and SBRT plan pretreatment quality assurance. Med Phys. 2020; 47(7):3153-3164. DOI: 10.1002/mp.14153. View

3.
Capaldi D, Skinner L, Dubrowski P, Yu A . An integrated quality assurance phantom for frameless single-isocenter multitarget stereotactic radiosurgery. Phys Med Biol. 2020; 65(11):115006. DOI: 10.1088/1361-6560/ab8534. View

4.
Nakano H, Tanabe S, Yamada T, Utsunomiya S, Takizawa T, Sakai M . Maximum distance in single-isocenter technique of stereotactic radiosurgery with rotational error using margin-based analysis. Radiol Phys Technol. 2021; 14(1):57-63. DOI: 10.1007/s12194-020-00602-2. View

5.
Parent L, Fielding A, Dance D, Seco J, Evans P . Amorphous silicon EPID calibration for dosimetric applications: comparison of a method based on Monte Carlo prediction of response with existing techniques. Phys Med Biol. 2007; 52(12):3351-68. DOI: 10.1088/0031-9155/52/12/003. View